Bloodborne Viral Pathogen Contamination in the Era of Laboratory Automation

Author:

Bryan Andrew1,Cook Linda1,Atienza Ederlyn E1,Kuypers Jane1,Cent Anne1,Baird Geoffrey S1,Coombs Robert W12,Jerome Keith R13,Wener Mark H12,Butler-Wu Susan M4

Affiliation:

1. Department of Laboratory Medicine and

2. Department of Medicine, University of Washington, Seattle, WA

3. Fred Hutchinson Cancer Research Center, Seattle, WA

4. Department of Pathology and Laboratory Medicine, Keck School of Medicine, University of Southern California, Los Angeles, CA

Abstract

Abstract BACKGROUND The CDC states that laboratory testing for persons under investigation for Ebola virus disease can be safely performed using automated laboratory instruments by adhering to bloodborne pathogen practices. We therefore sought to investigate the levels of viral contamination of a total laboratory automation (TLA) system to guide risk mitigation strategies for handling infectious agents. METHODS Environmental swabs followed by PCR for hepatitis B (HBV) and hepatitis C (HCV) viruses were taken from a chemistry TLA system during routine clinical use and after running a small number of high-titer HCV samples. Control experiments were performed to ensure the recovery of DNA and RNA viruses by swabs from a representative nonporous surface. RESULTS Of 79 baseline swabs for nucleic acids performed on the TLA system, 10 were positive for HBV and 8 for HCV. Viral nucleic acid was consistently detected from swabs taken from the distal inside surface of the decapper discharge chute, with areas adjacent to the decapper instrument and the centrifuge rotor also positive for HBV or HCV nucleic acid. Contamination was occasionally detected on exposed surfaces in areas without protective barriers between samples and personnel. After running known HCV-positive samples, at least one additional site of contamination was detected on an exposed area of the line. CONCLUSIONS A low level of viral contamination of automated clinical laboratory equipment occurs in clinical use. Given the risks associated with highly infectious agents, there is a need for risk-mitigation procedures when handling all samples.

Publisher

Oxford University Press (OUP)

Subject

Biochemistry (medical),Clinical Biochemistry

Reference36 articles.

1. CDC. Guidance for U.S. laboratories for managing and testing routine clinical specimens when there is a concern about Ebola virus disease. http://www.cdc.gov/vhf/ebola/healthcare-us/laboratories/safe-specimen-management.html (Accessed February 2016).

2. Laboratory automation systems and workcells. CAP Today. 2014. http://www.captodayonline.com/productguides/instruments/laboratory-automation-systems-and-workcells-august-2015.html (Accessed May 2016).

3. Notes from the field: occupationally acquired HIV infection among health care workers - United States, 1985–2013;Joyce;MMWR Morb Mortal Wkly Rep,2015

4. Laboratory-acquired infections;Singh;Clin Infect Dis,2009

5. A case of Ebola virus infection;Emond;Br Med J,1977

Cited by 17 articles. 订阅此论文施引文献 订阅此论文施引文献,注册后可以免费订阅5篇论文的施引文献,订阅后可以查看论文全部施引文献

同舟云学术

1.学者识别学者识别

2.学术分析学术分析

3.人才评估人才评估

"同舟云学术"是以全球学者为主线,采集、加工和组织学术论文而形成的新型学术文献查询和分析系统,可以对全球学者进行文献检索和人才价值评估。用户可以通过关注某些学科领域的顶尖人物而持续追踪该领域的学科进展和研究前沿。经过近期的数据扩容,当前同舟云学术共收录了国内外主流学术期刊6万余种,收集的期刊论文及会议论文总量共计约1.5亿篇,并以每天添加12000余篇中外论文的速度递增。我们也可以为用户提供个性化、定制化的学者数据。欢迎来电咨询!咨询电话:010-8811{复制后删除}0370