Monitoring Temporary Immunodepression by Flow Cytometric Measurement of Monocytic HLA-DR Expression: A Multicenter Standardized Study

Author:

Döcke Wolf-Dietrich1,Höflich Conny1,Davis Kenneth A2,Röttgers Karsten1,Meisel Christian1,Kiefer Paul3,Weber Stefan U4,Hedwig-Geissing Monika5,Kreuzfelder Ernst6,Tschentscher Peter7,Nebe Thomas8,Engel Andrea9,Monneret Guillaume10,Spittler Andreas11,Schmolke Kathrin1,Reinke Petra12,Volk Hans-Dieter1,Kunz Dagmar13

Affiliation:

1. Institute of Medical Immunology, Charité-Campus Mitte, Humboldt University of Berlin, and University Hospital, Charité Berlin, Berlin, Germany

2. Becton Dickinson, San Diego, CA

3. Institute of Clinical Chemistry and Pathobiochemistry, University Hospital, Aachen, Germany

4. Department of Aneasthesiology and Intensive Care Medicine, University Hospital, Bonn, Germany

5. Anaesthesia Clinic, University Hospital, Erlangen, Germany

6. Institute of Immunology, University Hospital, Essen, Germany

7. Institute of Clinical Chemistry, University Hospital, Hamburg-Eppendorf, Germany

8. Laboratory of Flowcytometry, University Hospital, Heidelberg, Germany

9. BD Biosciences, Heidelberg, Germany

10. Immunology Laboratory, Hopital Neurologique, Lyon South Hospital, Lyon, France

11. Surgical Research Laboratories, University of Vienna, Vienna, Austria

12. Department of Nephrology and Internal Intensive Care, Clinic of Internal Medicine, Charité-Campus Virchow, Humboldt University of Berlin, Berlin, Germany

13. Deutsches Rotes Kreuz Hospital, Neuwied, Germany

Abstract

Abstract Background: Single-center trials have shown that monocytic HLA-DR is a good marker for monitoring the severity of temporary immunodepression after trauma, major surgery, or sepsis. A new test for measuring monocytic HLA-DR is now available. Methods: We evaluated a new test reagent set for monocytic HLA-DR expression (BD Quantibrite™ HLA-DR/Monocyte reagent; Becton Dickinson) in single-laboratory and interlaboratory experiments, assessing preanalytical handling, lyse-no-wash (LNW) vs lyse-wash (LW) values, reference values, and the effect of use of different flow cytometers and different instrument settings on test variance. Results: For preanalytical handling, EDTA anticoagulation, storage on ice as soon as possible, and staining within 4 h after blood collection gave results comparable to values obtained for samples analyzed immediately after collection (mean increase of ∼4% in monocytic HLA-DR). Comparison of LNW and LW revealed slightly higher results for LNW (∼18% higher for LNW compared with LW; r = 0.982). Comparison of different flow cytometers and instrument settings gave CVs <4%, demonstrating the independence of the test from these variables and suggesting that this method qualifies as a standardized test. CV values from the interlaboratory comparison ranged from 15% (blood sample unprocessed before transport) to 25% (stained and fixed before transport). Conclusions: For the BD Quantibrite HLA-DR/Monocyte test, preanalytical handling is standardized. Single-laboratory results demonstrated the independence of this test from flow cytometer and instrument settings. Interlaboratory results showed greater variance than single-laboratory values. This interlaboratory variance was partly attributable to the influence of transport and can be reduced by optimization of transport conditions.

Publisher

Oxford University Press (OUP)

Subject

Biochemistry, medical,Clinical Biochemistry

Cited by 206 articles. 订阅此论文施引文献 订阅此论文施引文献,注册后可以免费订阅5篇论文的施引文献,订阅后可以查看论文全部施引文献

同舟云学术

1.学者识别学者识别

2.学术分析学术分析

3.人才评估人才评估

"同舟云学术"是以全球学者为主线,采集、加工和组织学术论文而形成的新型学术文献查询和分析系统,可以对全球学者进行文献检索和人才价值评估。用户可以通过关注某些学科领域的顶尖人物而持续追踪该领域的学科进展和研究前沿。经过近期的数据扩容,当前同舟云学术共收录了国内外主流学术期刊6万余种,收集的期刊论文及会议论文总量共计约1.5亿篇,并以每天添加12000余篇中外论文的速度递增。我们也可以为用户提供个性化、定制化的学者数据。欢迎来电咨询!咨询电话:010-8811{复制后删除}0370

www.globalauthorid.com

TOP

Copyright © 2019-2024 北京同舟云网络信息技术有限公司
京公网安备11010802033243号  京ICP备18003416号-3