Analytical and Clinical Evaluation of the Bayer ADVIA Centaur Automated B-Type Natriuretic Peptide Assay in Patients with Heart Failure: A Multisite Study

Author:

Wu Alan H B1,Packer Milton2,Smith Andrew1,Bijou Rachel2,Fink Daniel2,Mair Johannes3,Wallentin Lars4,Johnston Nina5,Feldcamp Carolyn S6,Haverstick Doris M7,Ahnadi Charaf E8,Grant Andrew8,Despres Normand9,Bluestein Barry9,Ghani Farooq9

Affiliation:

1. Department of Pathology and Laboratory Medicine, Hartford Hospital, 80 Seymour St., Hartford, CT 06102

2. Departments of Medicine and Pathology, Columbia Presbyterian Medical Center, New York, NY

3. Department of Internal Medicine/Division of Cardiology, University of Innsbruck, Innsbruck, Austria

4. Uppsala Clinical Research Center and

5. Department of Cardiology, Cardiothoracic Center, Uppsala University Hospital, Uppsala, Sweden

6. Pathology Department, Henry Ford Hospital, Detroit, MI

7. Department of Pathology, University of Virginia, Charlottesville, VA

8. Centre for Research and Evaluation in Diagnostics, CRED, Sherbrooke, Quebec, Canada

9. Bayer Healthcare LLC, Diagnostics Division, Tarrytown, NY

Abstract

Abstract Background: B-Type natriuretic peptide (BNP) is released from the left ventricle of the heart into the circulation in response to ventricular stretching and volume overload. Increased BNP concentrations are associated with heart failure (HF). Methods: We evaluated the analytical and clinical performance of the Bayer ADVIA Centaur® BNP assay. Studies included precision, analytical correlation (against the Shionogi ShionoRIA™ and Biosite Triage® BNP assays), BNP results for blood collected in plastic tubes containing EDTA vs other collection tubes, high-dose hook effect, detection limits, and interferences. The clinical performance was tested on 2243 blood samples collected from 983 apparently healthy individuals, 538 patients with chronic disease but without HF (renal insufficiency, chronic obstructive pulmonary disease, diabetes, and hypertension), and 722 patients with HF (New York Heart Association classes I-IV). Results: The ADVIA Centaur assay had total imprecision (CV) of 3.4%, 2.9%, and 2.4% at BNP concentrations of 48, 461, and 1768 ng/L, respectively. The Passing–Bablok correlations to the ShionoRIA and Triage were as follows: ADVIA Centaur = 1.11(ShionoRIA) − 1.19 ng/L (r = 0.98); ADVIA Centaur = 0.78(Triage) + 5.89 ng/L (r = 0.92), respectively. Of the different blood collection tubes, only EDTA plastic tubes (with and without the barrier gel) were acceptable. The lower detection limit was 0.5 ng/L, and there were no interferences from common analytes, other neuropeptides, or unusual antibodies. BNP exhibited different reference intervals according to age and gender. BNP concentrations increased progressively as the severity of HF increased. Conclusions: The ADVIA Centaur is the first commercially available BNP assay for use on an automated immunochemistry platform. This assay has good analytical and clinical performance characteristics for diagnosing HF.

Publisher

Oxford University Press (OUP)

Subject

Biochemistry (medical),Clinical Biochemistry

Cited by 120 articles. 订阅此论文施引文献 订阅此论文施引文献,注册后可以免费订阅5篇论文的施引文献,订阅后可以查看论文全部施引文献

同舟云学术

1.学者识别学者识别

2.学术分析学术分析

3.人才评估人才评估

"同舟云学术"是以全球学者为主线,采集、加工和组织学术论文而形成的新型学术文献查询和分析系统,可以对全球学者进行文献检索和人才价值评估。用户可以通过关注某些学科领域的顶尖人物而持续追踪该领域的学科进展和研究前沿。经过近期的数据扩容,当前同舟云学术共收录了国内外主流学术期刊6万余种,收集的期刊论文及会议论文总量共计约1.5亿篇,并以每天添加12000余篇中外论文的速度递增。我们也可以为用户提供个性化、定制化的学者数据。欢迎来电咨询!咨询电话:010-8811{复制后删除}0370

www.globalauthorid.com

TOP

Copyright © 2019-2024 北京同舟云网络信息技术有限公司
京公网安备11010802033243号  京ICP备18003416号-3