Affiliation:
1. Key Laboratory of Analytical Chemistry (Chongqing), College of Chemistry and Chemical Engineering, Southwest University, Chongqing, People’s Republic of China
Abstract
Abstract
Background: Methods based on magnetic bead probes have been developed for immunoassay, but most involve complicated labeling or stripping procedures and are unsuitable for routine use.
Methods: We synthesized magnet core/shell NiFe2O4/SiO2 nanoparticles and fabricated an electrochemical magnetic controlled microfluidic device for the detection of 4 tumor markers. The immunoassay system consisted of 5 working electrodes and an Ag/AgCl reference electrode integrated on a glass substrate. Each working electrode contained a different antibody immobilized on the NiFe2O4/SiO2 nanoparticle surface and was capable of measuring a specific tumor marker using noncompetitive electrochemical immunoassay.
Results: Under optimal conditions, the multiplex immunoassay enabled the simultaneous detection of 4 tumor markers. The sensor detection limit was <0.5 μg/L (or <0.5 kunits/L) for most analytes. Intra- and interassay imprecisions (CVs) were <4.5% and 8.7% for analyte concentrations >5 μg/L (or >5 kunits/L), respectively. No nonspecific adsorption was observed during a series of procedures to detect target proteins, and electrochemical cross-talk (CV) between neighboring sites was <10%.
Conclusion: This immunoassay system offers promise for label-free, rapid, simple, cost-effective analysis of biological samples. Importantly, the chip-based immunosensor could be suitable for use in the mass production of miniaturized lab-on-a-chip devices and open new opportunities for protein diagnostics and biosecurity.
Funder
Postgraduate Science and Technology Innovation Program of Southwest China University
National Natural Sciences Foundation of China
Publisher
Oxford University Press (OUP)
Subject
Biochemistry (medical),Clinical Biochemistry
Cited by
136 articles.
订阅此论文施引文献
订阅此论文施引文献,注册后可以免费订阅5篇论文的施引文献,订阅后可以查看论文全部施引文献