Background: In recent years, a large number of studies have been published on the performance of glucose monitors. The quality of these reports is not known.
Methods: We searched the PubMed database for performance evaluations of handheld glucose monitors published from August 2002 to November 2006. Relevant articles were compared to 20 recommendations from the Standards for Reporting Diagnostic Accuracy (STARD) and 18 recommendations from the Clinical and Laboratory Standards Institute (CLSI).
Results: A total of 52 reports met our inclusion criteria and were reviewed. None (0%) of the reports conformed to all 38 STARD and CLSI recommendations. The range of compliance to these recommendations varied widely (median 53%; range 21%–84%). Only 1 study of the 52 reported following a CLSI recommendation for checking reference test results. Fewer than half (42%) of the reports contained STARD-recommended statements regarding how and when comparative measurements were performed.
Conclusions: None of the glucose monitor reports from our review conformed to all STARD and CLSI recommendations. Our finding that the average rate of compliance to recommendations was low suggests that many of the researchers did not follow published recommendations for study design, methodology, and reporting and that study quality and conclusions may have been affected. Future studies evaluating the performance of glucose monitoring systems should be carefully designed and follow published recommendations for methodological and reporting quality.
Publisher
Oxford University Press (OUP)
Subject
Biochemistry, medical,Clinical Biochemistry
Reference70 articles.
1. Standards of Medical Care in Diabetes. Diabetes Care2006;29(Suppl):S4-42.
2. Gutman S, Bernhardt P, Pinkos A, Moxey-Mims M, Knott T, Cooper J. Regulatory aspects of invasive glucose measurements. Diabetes Technol Ther2002;4:775-777.
3. Review Criteria Assessment of Portable Blood Glucose Monitoring In Vitro Diagnostic Devices Using Glucose Oxidase, Dehydrogenase, or Hexokinase Methodology. HHS Publication FDA-96-604
1996 US Department of Health and Human Services, Public Health Service, Food and Drug Administration, Center for Devices and Radiological Health, Office of Device Evaluation, Division of Clinical Laboratory Services Rockville, MD. .
4. Nichols JH. Interpreting method evaluations. Diabetes Technol Ther2002;4:623-625.
5. Finkielman JD, Oyen LJ, Afessa B. Agreement between bedside blood and plasma glucose measurement in the ICU setting. Chest2005;127:1749-1751.