Alternative Calibration Strategies for the Clinical Laboratory: Application to Nortriptyline Therapeutic Drug Monitoring

Author:

Olson Matthew T1,Breaud Autumn1,Harlan Robert1,Emezienna Nkechinyere1,Schools Sabitha1,Yergey Alfred L2,Clarke William1

Affiliation:

1. Department of Pathology, the Johns Hopkins Hospital, Baltimore, MD

2. National Institute of Child Health and Human Development, National Institutes of Health, Bethesda, MD

Abstract

BACKGROUND The addition of a calibration curve with every run is both time-consuming and expensive for clinical mass spectrometry assays. We present alternative calibration strategies that eliminate the need for a calibration curve except as required by laboratory regulations. METHODS We measured serum nortriptyline concentrations prospectively in 68 patients on 16 days over a 2-month period using a method employing calibration schemes that relied on the measurement of the response ratio (RR) corrected by the response factor (RF), i.e., a measurement of the RR for an equimolar mixture of the analyte and internal standard. Measurements were taken with contemporaneous RF (cRF) measurements as well as sporadic RF (sRF) measurements. The measurements with these alternative calibration schemes were compared against the clinical results obtained by interpolation on a calibration curve, and those differences were evaluated for analytical and clinical significance. RESULTS The differences between the values measured by cRF and sRF calibration and interpolation on a calibration curve were not significant (P = 0.088 and P = 0.091, respectively). Both the cRF- and sRF-based calibration results demonstrated a low mean bias against the calibration curve interpolations of 3.69% (95% CI, −15.8% to 23.2%) and 3.11% (95% CI, −16.4% to 22.6%), respectively. When these results were classified as subtherapeutic, therapeutic, or supratherapeutic, there was categorical agreement in 95.6% of the cRF calibration results and 94.1% of the sRF results. CONCLUSIONS cRF and sRF calibration in a clinically validated liquid chromatography–tandem mass spectrometry assay yields results that are analytically and clinically commensurate to those produced by interpolation with a calibration curve.

Funder

Thermo Fisher Scientific

Publisher

Oxford University Press (OUP)

Subject

Biochemistry, medical,Clinical Biochemistry

同舟云学术

1.学者识别学者识别

2.学术分析学术分析

3.人才评估人才评估

"同舟云学术"是以全球学者为主线,采集、加工和组织学术论文而形成的新型学术文献查询和分析系统,可以对全球学者进行文献检索和人才价值评估。用户可以通过关注某些学科领域的顶尖人物而持续追踪该领域的学科进展和研究前沿。经过近期的数据扩容,当前同舟云学术共收录了国内外主流学术期刊6万余种,收集的期刊论文及会议论文总量共计约1.5亿篇,并以每天添加12000余篇中外论文的速度递增。我们也可以为用户提供个性化、定制化的学者数据。欢迎来电咨询!咨询电话:010-8811{复制后删除}0370

www.globalauthorid.com

TOP

Copyright © 2019-2024 北京同舟云网络信息技术有限公司
京公网安备11010802033243号  京ICP备18003416号-3