Evaluating Rapid Rule-out of Acute Myocardial Infarction Using a High-Sensitivity Cardiac Troponin I Assay at Presentation

Author:

Greenslade Jaimi12,Cho Elizabeth2,Van Hise Christopher2,Hawkins Tracey1,Parsonage William123,Ungerer Jacobus4,Tate Jillian4,Pretorius Carel4,Than Martin5,Cullen Louise123

Affiliation:

1. Royal Brisbane and Women's Hospital, Herston, Australia

2. Faculty of Medicine, The University of Queensland, Brisbane, Australia

3. Faculty of Health, Queensland University of Technology, Brisbane, Australia

4. Pathology Queensland, Herston, Australia

5. Christchurch Hospital, Christchurch, New Zealand

Abstract

Abstract BACKGROUND Low concentrations of cardiac troponin (cTn) have been recommended for rapid rule-out of acute myocardial infarction (AMI). We examined the Beckman Coulter Access high-sensitivity cardiac troponin I (hs-cTnI) assay to identify a single test threshold that can safely rule out AMI. METHODS This analysis used stored samples collected in 2 prospective observational studies. In all, 1871 patients presenting to a tertiary emergency department with symptoms of acute coronary syndrome had blood taken for measurement of cTnI on presentation. The endpoint was type 1 myocardial infarction (T1MI). Sensitivity and negative predictive value (NPV) were calculated for hs-cTnI values below the 99th percentile. RESULTS Ninety-eight patients had T1MI (5.2%), and 638 (34.1%) patients had an hs-cTnI <2 ng/L (limit of detection), with sensitivity of 99.0% (95% CI, 94.4%–100%) and NPV of 99.8% (95% CI, 99.1%–100%). No hs-cTnI value above a concentration of 2 ng/L achieved sensitivity of 99%. However, an NPV of 99.5% was achieved at values <6 ng/L. A cutoff <6 ng/L enabled 1475 (78.8%) patients to be ruled out on presentation with sensitivity of 93.9% (95% CI, 87.1%–97.7%). CONCLUSIONS A single baseline cTn <2 ng/L measured with the Access hs-cTnI assay performed well for rule-out of AMI. This cutoff concentration identified 99% of patients with AMI and could reduce the number of patients requiring lengthy assessment. A cutoff of <6 ng/L yielded a high NPV but missed more cases of AMI than would be acceptable to clinicians.

Publisher

Oxford University Press (OUP)

Subject

Biochemistry (medical),Clinical Biochemistry

同舟云学术

1.学者识别学者识别

2.学术分析学术分析

3.人才评估人才评估

"同舟云学术"是以全球学者为主线,采集、加工和组织学术论文而形成的新型学术文献查询和分析系统,可以对全球学者进行文献检索和人才价值评估。用户可以通过关注某些学科领域的顶尖人物而持续追踪该领域的学科进展和研究前沿。经过近期的数据扩容,当前同舟云学术共收录了国内外主流学术期刊6万余种,收集的期刊论文及会议论文总量共计约1.5亿篇,并以每天添加12000余篇中外论文的速度递增。我们也可以为用户提供个性化、定制化的学者数据。欢迎来电咨询!咨询电话:010-8811{复制后删除}0370

www.globalauthorid.com

TOP

Copyright © 2019-2024 北京同舟云网络信息技术有限公司
京公网安备11010802033243号  京ICP备18003416号-3