Isotope Dilution–LC-MS/MS Reference Method Assessment of Serum Folate Assay Accuracy and Proficiency Testing Consensus Mean

Author:

Blackmore Sheena1,Pfeiffer Christine M2,Lee Anne1,Fazili Zia2,Hamilton Malcolm S1

Affiliation:

1. UK NEQAS Haematinics Scheme for Haematinics, Good Hope Hospital, Sutton Coldfield, UK

2. Division of Laboratory Sciences, National Center for Environmental Health, Centers for Disease Control and Prevention, Atlanta, GA

Abstract

BACKGROUND Current methods for measuring folates in clinical laboratories are competitive folate binding protein assays. These assays show a considerable lack of agreement that has implications for the comparability of data across studies as well as for long-term population studies. The development of isotope dilution–liquid chromatography–tandem mass spectrometry (ID-LC-MS/MS) reference methods permitted the evaluation of method accuracy and consistency over time. METHODS We measured 3 pools of human serum by ID-LC-MS/MS, calculated values for total folate, and distributed the same pools to participants in a national External Quality Assessment scheme. We used linear regression to compare the all-laboratory and method data with reference method values. The exercise was repeated after 18 months to assess the stability of the all-laboratory and method data. RESULTS The distributed serum pools had mass spectrometry values for folate species typical of those found in healthy individuals from populations not receiving dietary folic acid fortification. There was good agreement of the all-laboratory data set with the reference method (y =0.86x + 0.91 μg/L) at both time points. Linear regression demonstrated that the Abbott Architect showed the closest agreement with the reference method. The Roche Elecsys method was nonlinear and showed a calibration offset of 2.6 μg/L (4.57 nmol/L). CONCLUSIONS Calibration of serum folate assays traceable to higher-order reference methods increases method accuracy and improves consistency. The all-laboratory consensus mean proved sufficiently accurate and stable to be used as the target for monitoring laboratory performance.

Publisher

Oxford University Press (OUP)

Subject

Biochemistry, medical,Clinical Biochemistry

Reference29 articles.

1. Accuracy in clinical chemistry: who will kiss Sleeping Beauty awake?;Thienpont;Clin Chem Lab Med,2008

2. Traceability in laboratory medicine;Vesper;Clin Chem,2009

3. How far can the concepts of traceability and GUM/VIM can be applied to measurement results in laboratory medicine?;Goldschmidt;Accred Qual Assur,2004

4. Application of traceability concepts to analytical quality control may reconcile total error with uncertainty of measurement;Panteghini;Clin Chem Lab Med,2010

5. Folate supplements prevent recurrence of neural tube defects. FDA Dietary Supplement Task Force;Rush;Nutrition Rev,1992

Cited by 19 articles. 订阅此论文施引文献 订阅此论文施引文献,注册后可以免费订阅5篇论文的施引文献,订阅后可以查看论文全部施引文献

同舟云学术

1.学者识别学者识别

2.学术分析学术分析

3.人才评估人才评估

"同舟云学术"是以全球学者为主线,采集、加工和组织学术论文而形成的新型学术文献查询和分析系统,可以对全球学者进行文献检索和人才价值评估。用户可以通过关注某些学科领域的顶尖人物而持续追踪该领域的学科进展和研究前沿。经过近期的数据扩容,当前同舟云学术共收录了国内外主流学术期刊6万余种,收集的期刊论文及会议论文总量共计约1.5亿篇,并以每天添加12000余篇中外论文的速度递增。我们也可以为用户提供个性化、定制化的学者数据。欢迎来电咨询!咨询电话:010-8811{复制后删除}0370

www.globalauthorid.com

TOP

Copyright © 2019-2024 北京同舟云网络信息技术有限公司
京公网安备11010802033243号  京ICP备18003416号-3