Inaccurate First-Generation Testosterone Assays Are Influenced by Sex Hormone–Binding Globulin Concentrations

Author:

Heijboer Annemieke C1,Savelkoul Edo1,Kruit Adrian2,Endert Erik3,Blankenstein Marinus A1

Affiliation:

1. Department of Clinical Chemistry, Endocrine Laboratory, VU University Medical Center, Amsterdam, the Netherlands

2. Medical Laboratory, Nij Smellinghe Hospital, Drachten, the Netherlands

3. Department of Clinical Chemistry, Laboratory of Endocrinology, Academic Medical Center, University of Amsterdam, Amsterdam, the Netherlands

Abstract

Abstract Background The quality of testosterone assays has been a matter of debate for several years. Known limitations of testosterone immunoassays are the cross-reactivity with other steroids and a high variation in the low concentration range. We hypothesized that one of the additional limitations of testosterone immunoassays is an ineffective displacement of testosterone from its binding protein. Methods Thirty samples from women not using oral contraceptives (OAC), 30 samples from women using OAC, and 30 samples from pregnant women were used to measure testosterone by an isotope dilution (ID)-LC-MS/MS method and by 6 commercially available testosterone immunoassays (UniCel®, ARCHITECT®, Centaur®, Cobas®, Immulite®, and Liaison®). In addition, sex hormone–binding globulin (SHBG)4 was measured by immunoassay (ARCHITECT). Results The first-generation immunoassays (UniCel, Centaur, Immulite, and Liaison) showed inaccurate testosterone results in the method comparisons with the ID-LC-MS/MS method (R between 0.61 and 0.86) and for some assays (UniCel and Liaison) also a very poor standardization (slopes of 0.59 and 0.67, respectively). On average, SHBG concentrations were lowest in women not using OAC and highest in pregnant women, and overall ranged from 18.5 to 633 nmol/L. In the first-generation immunoassays, but not in the second-generation immunoassays, we observed an inverse relationship between SHBG concentrations and deviations in testosterone from the ID-LC-MS/MS results. Conclusions Widely used first-generation testosterone immunoassays are influenced by SHBG concentrations, which lead to inaccurate results in samples from patients with high or low SHBG concentrations, respectively. Laboratory specialists, clinicians, and researchers should be aware of this limitation in testosterone assays.

Publisher

Oxford University Press (OUP)

Subject

General Medicine

Cited by 14 articles. 订阅此论文施引文献 订阅此论文施引文献,注册后可以免费订阅5篇论文的施引文献,订阅后可以查看论文全部施引文献

同舟云学术

1.学者识别学者识别

2.学术分析学术分析

3.人才评估人才评估

"同舟云学术"是以全球学者为主线,采集、加工和组织学术论文而形成的新型学术文献查询和分析系统,可以对全球学者进行文献检索和人才价值评估。用户可以通过关注某些学科领域的顶尖人物而持续追踪该领域的学科进展和研究前沿。经过近期的数据扩容,当前同舟云学术共收录了国内外主流学术期刊6万余种,收集的期刊论文及会议论文总量共计约1.5亿篇,并以每天添加12000余篇中外论文的速度递增。我们也可以为用户提供个性化、定制化的学者数据。欢迎来电咨询!咨询电话:010-8811{复制后删除}0370

www.globalauthorid.com

TOP

Copyright © 2019-2024 北京同舟云网络信息技术有限公司
京公网安备11010802033243号  京ICP备18003416号-3