Stability of Biochemical Components in Blood Samples Transported by Tempus600/Sysmex GLP Robot Reception System

Author:

Andersen Ida Boegh1,Mogensen Nina1,Brandslund Ivan12

Affiliation:

1. Department of Clinical Immunology and Biochemistry, Lillebaelt Hospital, Vejle, Denmark

2. Faculty of Health Science, University of Southern Denmark, Odense, Denmark

Abstract

Abstract Background The Department of Clinical Immunology and Biochemistry, Lillebaelt Hospital, Denmark, and TIMEDICO invented and developed the Tempus600®, a pneumatic tube system (PTS). This system is only 2.5 cm in diameter and sends blood tubes directly, without packing before transport. Request forms need not be enclosed, since samples are labeled with patient and barcode ID. At the end of the pipes, a robot receives and places the tubes onto a GLP transport system after the principle of “first in, first out” (FIFO), whereafter the tubes are delivered to centrifuges and analytical equipment. The whole system has decreased the total turnaround time (ToTAT) from 2 to 3 h to <60 min. The aim of the study was to validate the Tempus600/GLP robot system for the 89 most frequent or critical components. Methods Two sets of blood samples were drawn from 100 patients. One set was sent by the Tempus600/GLP robot system, and the other was couriered directly to the laboratory for analysis. Results were compared using difference plots. Results The results for 85 of the components did not deviate from the optimized fast courier reference transport more than could be expected by the analytical variation. O2 saturation, oxyhemoglobin, and partial pressure of oxygen (P  o  2) deviated considerably from a clinical point of view. Partial pressure of carbon dioxide (P  co  2) results were clinically acceptable for assessing patients with obstructive lung disease for hypercapnia. Conclusions The Tempus600/GLP robot system can be used for transport of the majority of routinely used analytical tests.

Publisher

Oxford University Press (OUP)

Subject

General Medicine

Reference9 articles.

1. Stability of biochemical components in blood samples transported by the new dedicated blood tube transport system, Tempus 600 [in Danish];Hastrup,2013

2. Collection of diagnostic venous blood specimens;CLSI,2016

3. Statistical methods for assessing agreement between two methods of clinical measurement;Bland;Lancet,1986

4. Drwiega  J, Chaffin  C, Hardy  R. Adventures in turnaround time. AACC; 2016. https://www.aacc.org/publications/cln/articles/2016/june/adventures-in-turnaround-time (Accessed June 2016).

5. Kirkwood  J . The next step for molecular point-of-care testing: shrinking technology and tighter turnaround times shake up the field. AACC; 2016. https://www.aacc.org/publications/cln/articles/2016/june/the-next-step-for-molecular-point-of-care-testing (Accessed June 2016).

Cited by 16 articles. 订阅此论文施引文献 订阅此论文施引文献,注册后可以免费订阅5篇论文的施引文献,订阅后可以查看论文全部施引文献

同舟云学术

1.学者识别学者识别

2.学术分析学术分析

3.人才评估人才评估

"同舟云学术"是以全球学者为主线,采集、加工和组织学术论文而形成的新型学术文献查询和分析系统,可以对全球学者进行文献检索和人才价值评估。用户可以通过关注某些学科领域的顶尖人物而持续追踪该领域的学科进展和研究前沿。经过近期的数据扩容,当前同舟云学术共收录了国内外主流学术期刊6万余种,收集的期刊论文及会议论文总量共计约1.5亿篇,并以每天添加12000余篇中外论文的速度递增。我们也可以为用户提供个性化、定制化的学者数据。欢迎来电咨询!咨询电话:010-8811{复制后删除}0370

www.globalauthorid.com

TOP

Copyright © 2019-2024 北京同舟云网络信息技术有限公司
京公网安备11010802033243号  京ICP备18003416号-3