Affiliation:
1. Madeline Smith and Marya Lieberman, University of Notre Dame, Notre Dame, IN; and Ayenew Ashenef, Addis Ababa University, Addis Ababa, Ethiopia.
Abstract
A paper analytic device, the chemoPAD, was developed and validated to visually detect methotrexate, doxorubicin, cisplatin, and oxaliplatin at concentrations commonly found in injectable dosage forms. By testing residual solution after patient treatment, the chemoPAD can be used to monitor drug quality without restriction of patient access to medication. The chemoPAD is produced by wax printing on Ahlstrom paper to create separate reaction areas and deposits small amounts of chemicals to create color changes in response to different active pharmaceutical ingredients (APIs). This creates a unique color bar code to identify each medication. Internal validation studies show that the chemoPAD has excellent sensitivity and specificity to differentiate between samples of 100% and 0% API, which is the distinction relevant to the majority of reported falsified chemotherapy cases. The platinum-containing drugs, cisplatin and oxaliplatin, can be detected semiquantitatively. The cards can be read either visually by comparison with a standard image or by using computer image analysis. Dosage forms were collected from the Ethiopian health care system and analyzed with the chemoPAD followed by high-performance liquid chromatography. A substandard sample was discovered and reported to the Ethiopian Food Medicine and Healthcare Administration and Control Authority.
Publisher
American Society of Clinical Oncology (ASCO)
Cited by
4 articles.
订阅此论文施引文献
订阅此论文施引文献,注册后可以免费订阅5篇论文的施引文献,订阅后可以查看论文全部施引文献