Importance of Cross-Calibration When Replacing DXA Scanners: QDR4500W and Discovery Wi

Author:

Covey Margaret K.,Smith Donald L.,Berry Jean K.,Hacker Eileen D.

Abstract

The aim of the study was to determine reproducibility and extent of agreement between 2 dual-energy X-ray absorptiometers (Hologic QDR4500W, Discovery Wi). The average age of the sample (n = 42) was 50.4 (SD = 9.9) years old and 27.1 (SD = 6.1) kg/m2 body mass index. Four scans were performed with each subject (2 on each device) over ∼3 weeks. Whole body, proximal femur, and spine scans were performed at each visit. Major variables were whole body bone mineral content (BMC), fat mass, and nonosseous lean mass, and bone mineral density (BMD) of total proximal femur, femoral neck, total spine. Bland and Altman plots assessed the extent of the agreement. Regression analysis was used to develop correction equations if indicated. Both devices demonstrated good precision for whole body composition and BMD of central sites (<1% different). Interdevice agreement was acceptable for BMD of central sites (<1% different), but there were systematic differences for whole body composition between the 2 devices. It was concluded that when replacing an existing scanner with a new model, in vivo cross-calibration is important to ensure comparability of scan data, especially for whole body composition.

Publisher

Springer Publishing Company

Subject

General Medicine,General Nursing

同舟云学术

1.学者识别学者识别

2.学术分析学术分析

3.人才评估人才评估

"同舟云学术"是以全球学者为主线,采集、加工和组织学术论文而形成的新型学术文献查询和分析系统,可以对全球学者进行文献检索和人才价值评估。用户可以通过关注某些学科领域的顶尖人物而持续追踪该领域的学科进展和研究前沿。经过近期的数据扩容,当前同舟云学术共收录了国内外主流学术期刊6万余种,收集的期刊论文及会议论文总量共计约1.5亿篇,并以每天添加12000余篇中外论文的速度递增。我们也可以为用户提供个性化、定制化的学者数据。欢迎来电咨询!咨询电话:010-8811{复制后删除}0370

www.globalauthorid.com

TOP

Copyright © 2019-2024 北京同舟云网络信息技术有限公司
京公网安备11010802033243号  京ICP备18003416号-3