CALCULATION OF DEFORMATIONS AND EVALUATION OF STATIC STRENGTH OF STRUCTURAL ELEMENTS OF IMPLANT SYSTEMS ON THE EXAMPLE OF DENTAL IMPLANTS LIKO-M 4×10 OF DIFFERENT DESIGN

Author:

Sarkisov Armen1,Zelensky Vladimir2,Avanisyan Vazgen2,Dolgalev Aleksandr23,Sergeev Yuriy2,Semerikov Dmitriy4,Spitsyna Anna1

Affiliation:

1. Astrakhan State Medical University

2. Stavropol State Medical university

3. Pyatigorsk Medical and Pharmaceutical Institute - a branch of the Volgograd State Medical University

4. Valentina Dental Clinic

Abstract

Introduction. The creation of effective, reliable, safe, technologically advanced and competitive products is the main task of medical device engineering. The most important requirement to modern medical devices is to guarantee patients' safety during their lifetime. Today the use of modern computer-aided engineering analysis packages is the most effective calculation method for evaluating the strength and reliability of unique medical devices that can lead to serious consequences if their operation is disturbed. One of the most suitable and efficient systems for automated engineering calculations (CAE-system) is the ANSYS software. The purpose of this study was the comparative assessment of the elastic and elastoplastic deformation of 4×10 dental implants of different designs on the abutment-pin and screw-body interface, using the computer simulation of the stress-strain state. Materials and methods. Two kinds of dental implants were chosen for this study: a Liko-M 4×10 implant with the cylindrical body shape and a Liko-M DG 4×10 implant with a tapered body shape. The contact between the abutment and screw as well as the implant body and screw is frictional. The pre-tensioning of the screw from the initial tightening was 400 N. The load was applied to the cylindrical surface of the abutment at a percentage of its height. Results. Elastic and elastic-plastic calculations of the stress-strain state of Liko-M 4×10 and Liko-M DG 4×10 implants were performed. Besides the results of the main calculations of the stress-strain state of the implants Liko-M 4×10 and Liko-M DG 4×10, necessary to assess their static strength, we have also calculated the strength coefficients of implant bodies. Comparative analysis of the static strength of the Liko-M 4×10 and Liko-M DG 4×10 implants provides conclusions, which are significant for practical application of the implants.

Publisher

Astrakhan State Medical University

Reference3 articles.

1. Пахарьков Г. Н., Попечителев Е. П. Принципы и методы обеспечения качества медико-технического оснащения здравоохранения. Учебное пособие. СПб.: СПб ГЭТУ «ЛЭТИ», 2003. 120 с., Pakhar'kov G. N., Popechitelev E. P. Principles and methods for ensuring the quality of medical and technical equipment for health care: a textbook. Saint Petersburg: Saint Petersburg Electrotechnical University “LETI”; 2003. 120 p.

2. Theory Reference for ANSYS and Workbench. Canonsburg: ANSYS Inc., 2019., Theory Reference for ANSYS and Workbench. Canonsburg: ANSYS Inc.; 2019.

3. Structural Analysis Guide. Canonsburg: ANSYS Inc., 2019, Structural Analysis Guide. Canonsburg: ANSYS Inc.; 2019.

Cited by 1 articles. 订阅此论文施引文献 订阅此论文施引文献,注册后可以免费订阅5篇论文的施引文献,订阅后可以查看论文全部施引文献

同舟云学术

1.学者识别学者识别

2.学术分析学术分析

3.人才评估人才评估

"同舟云学术"是以全球学者为主线,采集、加工和组织学术论文而形成的新型学术文献查询和分析系统,可以对全球学者进行文献检索和人才价值评估。用户可以通过关注某些学科领域的顶尖人物而持续追踪该领域的学科进展和研究前沿。经过近期的数据扩容,当前同舟云学术共收录了国内外主流学术期刊6万余种,收集的期刊论文及会议论文总量共计约1.5亿篇,并以每天添加12000余篇中外论文的速度递增。我们也可以为用户提供个性化、定制化的学者数据。欢迎来电咨询!咨询电话:010-8811{复制后删除}0370

www.globalauthorid.com

TOP

Copyright © 2019-2024 北京同舟云网络信息技术有限公司
京公网安备11010802033243号  京ICP备18003416号-3