Abstract
The resolution function of a spectrometer based on a strongly bent single crystal (bending radius of 10 cm or less) is evaluated. It is shown that the resolution is controlled by two parameters: (i) the ratio of the lattice spacing of the chosen reflection to the crystal thickness and (ii) a single parameter comprising crystal thickness, its bending radius, distance to a detector, and anisotropic elastic constants of the chosen crystal. The results allow the optimization of the parameters of bent-crystal spectrometers for the hard X-ray free-electron laser sources.