Abstract
In the last two decades, far-field high-energy diffraction microscopy (FF-HEDM) and similar non-destructive techniques have been actively developed at synchrotron light sources around the world. As these techniques (and associated analysis tools) are becoming more available for the general users of these light sources, it is important and timely to characterize their performance and capabilities. In this work, the FF-HEDM instrument implemented at the 1-ID-E endstation of the Advanced Photon Source (APS) is summarized. The set of measurements conducted to characterize the instrument's repeatability and sensitivity to changes in grain orientation and position are also described. When an appropriate grain matching method is used, the FF-HEDM instrument's repeatability is approximately 5 µm in translation, 0.02° in rotation, and 2 × 10−4 in strain; the instrument sensitivity is approximately 5 µm in translation and 0.05° in rotation.
Funder
US Department of Energy, Office of Science
Publisher
International Union of Crystallography (IUCr)
Subject
Instrumentation,Nuclear and High Energy Physics,Radiation
Cited by