Abstract
The Hard X-ray Photo-Electron Spectroscopy (HAXPES) beamline (PES-BL14), installed at the 1.5 T bending-magnet port at the Indian synchrotron (Indus-2), is now available to users. The beamline can be used for X-ray photo-emission electron spectroscopy measurements on solid samples. The PES beamline has an excitation energy range from 3 keV to 15 keV for increased bulk sensitivity. An in-house-developed double-crystal monochromator [Si (111)] and a platinum-coated X-ray mirror are used for the beam monochromatization and manipulation, respectively. This beamline is equipped with a high-energy (up to 15 keV) high-resolution (meV) hemispherical analyzer with a microchannel plate and CCD detector system with SpecsLab Prodigy and CasaXPS software. Additional user facilities include a thin-film laboratory for sample preparation and a workstation for on-site data processing. In this article, the design details of the beamline, other facilities and some recent scientific results are described.
Publisher
International Union of Crystallography (IUCr)
Subject
Instrumentation,Nuclear and High Energy Physics,Radiation
Cited by
42 articles.
订阅此论文施引文献
订阅此论文施引文献,注册后可以免费订阅5篇论文的施引文献,订阅后可以查看论文全部施引文献