Author:
Wernecke Jan,Gollwitzer Christian,Müller Peter,Krumrey Michael
Abstract
A dedicated in-vacuum X-ray detector based on the hybrid pixel PILATUS 1M detector has been installed at the four-crystal monochromator beamline of the PTB at the electron storage ring BESSY II in Berlin, Germany. Owing to its windowless operation, the detector can be used in the entire photon energy range of the beamline from 10 keV down to 1.75 keV for small-angle X-ray scattering (SAXS) experiments and anomalous SAXS at absorption edges of light elements. The radiometric and geometric properties of the detector such as quantum efficiency, pixel pitch and module alignment have been determined with low uncertainties. The first grazing-incidence SAXS results demonstrate the superior resolution in momentum transfer achievable at low photon energies.
Publisher
International Union of Crystallography (IUCr)
Subject
Instrumentation,Nuclear and High Energy Physics,Radiation
Reference30 articles.
1. A quarter-century of metrology using synchrotron radiation by PTB in Berlin
2. JCPDS—International Centre for Diffraction Data round robin study of silver behenate. A possible low-angle X-ray diffraction calibration standard
3. The PILATUS 1M detector
4. Chmeissani, M., Maiorino, M., Blanchot, G., Pellegrini, G., Garcia, J., Lozano, M., Martinez, R., Puigdengoles, C. & Mullan, M. (2004). Proceedings of the 21st IEEE Instrumentation and Measurement Technology Conference (IMTC 04), Vol. 1, pp. 787-791.
5. XPAD: A photons counting pixel detector for material sciences and small-animal imaging
Cited by
67 articles.
订阅此论文施引文献
订阅此论文施引文献,注册后可以免费订阅5篇论文的施引文献,订阅后可以查看论文全部施引文献