Abstract
Laue 3D neutron diffraction tomography (Laue 3DNDT) is a novel tomographic approach that enables non-destructive resolution of the 3D microstructure of polycrystalline materials, utilizing a high-intensity white beam of neutrons, resulting in high experimental efficiency. This article describes the functionality of the fast-acquisition Laue camera for neutrons (FALCON), a double-detector system setup used for Laue 3DNDT experiments at the pulse-overlap diffractometer (POLDI) at the SINQ neutron source of the Paul Scherrer Institute. The reported commissioning includes standard experimental protocols, strategies for acquiring data and a detailed characterization of the measurement system. Finally, single-crystal measurements for calibration and a full grain map of a CoNiGa oligocrystalline sample, extracted from a pilot study, are presented. A detailed understanding of the performance of the FALCON instrumentation is vital for the future advancement of the Laue 3DNDT algorithm, including, but not limited to, improved morphology reconstructions and strain mapping.
Funder
DanScatt
Board of the Swiss Federal Institutes of Technology
Ministerstvo Školství, Mládeže a Tělovýchovy
Publisher
International Union of Crystallography (IUCr)
Subject
General Biochemistry, Genetics and Molecular Biology
Cited by
1 articles.
订阅此论文施引文献
订阅此论文施引文献,注册后可以免费订阅5篇论文的施引文献,订阅后可以查看论文全部施引文献