Abstract
X-ray atomic pair distribution function (XPDF) measurements using a two-dimensional area detector have been of great value in studying atomic structures of materials with varying degrees of disorder. However, an area detector does not have energy resolution. Thus, incoherent inelastic Compton scattering and fluorescence are not discriminated, contaminating coherent elastic scattering. This paper investigates the effects of random noise from incoherent scattering on XPDF analysis. To conduct the study, the elastic scattering, Compton scattering and fluorescence of In0.33Ga0.67As alloy were separately measured using an intrinsic Ge solid-state detector with energy resolution. It is found that the addition of Compton scattering with a noise-to-signal ratio of about 0.8% results in the smearing of diffuse scattering in the high-Q region. Moreover, adding extra noise from fluorescence increases the smearing, overwhelming the diffuse scattering. Additionally, simulated data of elastic and Compton scattering on ferroelectric Ba(Ti0.8Zr0.2)O3 were used to investigate the evolution of noise fluctuation and its effects on the XPDF as a function of total intensity.
Funder
Pusan National University
Publisher
International Union of Crystallography (IUCr)
Subject
General Biochemistry, Genetics and Molecular Biology
Cited by
1 articles.
订阅此论文施引文献
订阅此论文施引文献,注册后可以免费订阅5篇论文的施引文献,订阅后可以查看论文全部施引文献