Abstract
The coverage of a given diffraction instrument as a percentage of the area 2π of a pole figure hemisphere is a crucial parameter of each diffraction instrument used for texture or strain pole figure determination. On the basis of this knowledge, the number of rotations and rotation angles for a full determination of the orientation distribution function can be optimized. However, the determination of this quantity is non-trivial. This paper presents a method that projects a given detector coverage into pole figure space, i.e. outlines the detector areas in a pole figure, and then determines the fraction of the entire 2π pole figure hemisphere around the sample that is covered. The freely available Generic Mapping Tools (GMT) and ImageJ are utilized for this quantification. With this method, it is shown that the empirically determined rotation angles for the HIPPO neutron time-of-flight diffractometer are close to optimal for a set of three rotations.
Funder
U.S. Department of Energy, National Nuclear Security Administration
Publisher
International Union of Crystallography (IUCr)
Subject
General Biochemistry, Genetics and Molecular Biology
Cited by
43 articles.
订阅此论文施引文献
订阅此论文施引文献,注册后可以免费订阅5篇论文的施引文献,订阅后可以查看论文全部施引文献