Scanning X-ray spectrometer for high-resolution Compton profile measurements at ESRF

Author:

Suortti P.,Buslaps T.,Fajardo P.,Honkimäki V.,Kretzschmer M.,Lienert U.,McCarthy J. E.,Renier M.,Shukla A.,Tschentscher Th.,Meinander T.

Abstract

A scanning-type crystal spectrometer for high-resolution Compton profile measurements has been constructed at the High Energy Inelastic Scattering Beamline (ID15B) of the ESRF. Radiation from a seven-period asymmetrical permanent-magnet wiggler or from a superconducting wavelength shifter is focused horizontally onto the sample by a bent-crystal monochromator. Typical energies are 30, 50 and 60 keV, the flux on the sample is 1012 photons s−1, and the relative energy bandwidth is 3 × 10−4. The spectrometer operates in the Rowland circle geometry, where the sample is fixed and the cylindrically bent analyser crystal and the detector move on the focusing circle by synchronized translations and rotations. The main detector is a large-diameter NaI scintillation counter, the incident beam is monitored by an Si diode, and scattering from the sample is detected using a Ge detector. The recorded spectrum is corrected for the energy-dependent response of the spectrometer, background and multiple scattering, and converted to the momentum scale. The resolution of the spectrometer is calculated from the geometrical factors and the reflectivity curve of the analyser crystal, and the result is checked against the widths of the elastically scattered line and fluorescent lines. So far, 0.1 a.u. resolution in electron momentum has been achieved. The typical average count rate over the Compton profile is about 1000 counts s−1 from a weakly absorbing sample.

Publisher

International Union of Crystallography (IUCr)

Subject

Instrumentation,Nuclear and High Energy Physics,Radiation

Cited by 61 articles. 订阅此论文施引文献 订阅此论文施引文献,注册后可以免费订阅5篇论文的施引文献,订阅后可以查看论文全部施引文献

1. High-resolution Compton spectroscopy using x-ray microcalorimeters;Review of Scientific Instruments;2022-11-01

2. X‐ray and Neutron Diffraction and Scattering of Clathrate Hydrates;Clathrate Hydrates;2022-02-11

3. An Overview of the Compton Scattering Calculation;Crystals;2021-05-10

4. The five-analyzer point-to-point scanning crystal spectrometer at ESRF ID26;Journal of Synchrotron Radiation;2021-01-01

5. High-Energy X-Ray Scattering and Imaging;Synchrotron Light Sources and Free-Electron Lasers;2020