Abstract
Absolute calibration of a soft X-ray spectrograph has been performed using a white beam of synchrotron radiation. The calibrated spectrograph was a flat-field grazing-incidence spectrograph with an X-ray CCD detector for X-ray laser research. Absolute sensitivity of the spectrograph system can be obtained from transmitted spectra using filters made of several different materials, each providing an absorption-edge wavelength standard. The absolute sensitivity determined in this work shows nearly the same behaviour with wavelength as that in another calibration experiment using a laser-produced plasma as an X-ray source.