Affiliation:
1. Al-Balqa Applied University
2. Philadelphia University
3. University of Jordan
Abstract
In this work, the utilization of the modified iodine-coated polycrystalline platinum electrode as a voltammetric sensor for copper determination in pharmaceutical formulations was reported. The optimized experimental parameters for the determination of copper were using 0.1M KCl as a supporting electrolyte with a scan rate of 50 mV/s, deposition potential of - 0.2 V, and a deposition time of 2 minutes. The anodic peak related to Copper oxidation is centered at about + 0.05 V. The extended detected linear range for the developed method was between 1 ppm and 100 ppm. The anodic current showed excellent linearity with R2 = 0.9986. The limit of detection (LOD) and limit of quantitation (LOQ) were 0.115 ppm and 0.346 ppm, respectively, which attests to the sensitivity of the method. The investigation for the effect of potential interferences from multivitamins tablet ingredients indicated a specific selectivity toward copper and the absence of any electrochemical response toward these components. The developed method was successfully applied to analyze copper and the obtained results were in good agreement with the labeled values, besides that, the statistical tests indicated no significant difference at p = 0.05 with 95 % confidence level.
Publisher
Taras Shevchenko National University of Kyiv
Cited by
6 articles.
订阅此论文施引文献
订阅此论文施引文献,注册后可以免费订阅5篇论文的施引文献,订阅后可以查看论文全部施引文献