Affiliation:
1. Cerebrovascular Disease Research Center, Department of Neurology, Cornell University Medical College, New York, New York, U.S.A.
Abstract
Laser-Doppler flowmetry (LDF), a new method allowing instantaneous, continuous, and noninvasive measurements of microcirculatory blood flow in a small tissue sample, was evaluated for its accuracy in monitoring regional cerebral blood flow (rCBF) in the cortical microcirculation after focal cerebral ischemia. Wistar and spontaneously hypertensive rats (SHR, n = 19) were subjected to permanent occlusion of the middle cerebral and common carotid arteries. Absolute rCBF in a tissue sample of the ischemic hemisphere was measured autoradiographically with [14C]iodoantipyrine as a tracer and compared to rCBF measured by LDF. Additionally, the percent change in rCBF between baseline and ischemic values was compared for both methods. Absolute rCBF values recorded with LDF correlated poorly ( r = 0.54) with [14C]iodoantipyrine measurements. In contrast, LDF readings expressed as a percentage of ischemic vs. preocclusion readings (relative LDF readings) correlated very well ( r = 0.91) with the percent change in [14C]iodoantipyrine measurements. We conclude that LDF does not provide accurate measurements of absolute rCBF values but this method allows accurate measurements of changes in rCBF due to induction of focal cerebral ischemia.
Subject
Cardiology and Cardiovascular Medicine,Clinical Neurology,Neurology
Cited by
469 articles.
订阅此论文施引文献
订阅此论文施引文献,注册后可以免费订阅5篇论文的施引文献,订阅后可以查看论文全部施引文献