Abstract
AbstractMagnetocardiography is a contactless imaging modality for electric current propagation in the cardiovascular system. Although conventional sensors provide sufficiently high sensitivity, their spatial resolution is limited to a centimetre-scale, which is inadequate for revealing the intra-cardiac electrodynamics such as rotational waves associated with ventricular arrhythmias. Here, we demonstrate invasive magnetocardiography of living rats at a millimetre-scale using a quantum sensor based on nitrogen-vacancy centres in diamond. The acquired magnetic images indicate that the cardiac signal source is well explained by vertically distributed current dipoles, pointing from the right atrium base via the Purkinje fibre bundle to the left ventricular apex. We also find that this observation is consistent with and complementary to an alternative picture of electric current density distribution calculated with a stream function method. Our technique will enable the study of the origin and progression of various cardiac arrhythmias, including flutter, fibrillation, and tachycardia.
Funder
MEXT | JST | Precursory Research for Embryonic Science and Technology
Ministry of Education, Culture, Sports, Science and Technology
Publisher
Springer Science and Business Media LLC
Subject
General Physics and Astronomy
Cited by
45 articles.