Varanges Vincent,Caglar Baris,Lebaupin Yann,Batt Till,He Weidong,Wang Jing,Rossi René M.,Richner Gilles,Delaloye Jean-Romain,Michaud Véronique
Abstract
AbstractAfter the spread of COVID-19, surgical masks became highly recommended to the public. They tend to be handled and used multiple times, which may impact their performance. To evaluate this risk, surgical masks of Type IIR were submitted to four simulated treatments: folding, ageing with artificial saliva or sweat and washing cycles. The air permeability, mechanical integrity, electrostatic potential, and filtration efficiency (FE) of the masks were measured to quantify possible degradation. Overall, air permeability and mechanical integrity were not affected, except after washing, which slightly degraded the filtering layers. Electrostatic potential and FE showed a strong correlation, highlighting the role of electrostatic charges on small particle filtration. A slight decrease in FE for 100 nm particles was found, from 74.4% for the reference masks to 70.6% for the mask treated in saliva for 8 h. A strong effect was observed for washed masks, resulting in FE of 46.9% (± 9.5%), comparable to that of a control group with no electrostatic charges. A dry store and reuse strategy could thus be envisaged for the public if safety in terms of viral and bacterial charge is ensured, whereas washing strongly impacts FE and is not recommended.
Funder
Innosuisse - Schweizerische Agentur für Innovationsförderung