Assessing mechanical catheter dysfunction in automated tidal peritoneal dialysis using cycler software: a case control, proof-of-concept study

Author:

Oviedo Flores KrystellORCID,Kaltenegger Lukas,Eibensteiner FabianORCID,Unterwurzacher Markus,Kratochwill KlausORCID,Aufricht ChristophORCID,König FranzORCID,Vychytil Andreas

Abstract

AbstractNew recommendations on evaluation of peritoneal membrane function suggest ruling out catheter dysfunction when evaluating patients with low ultrafiltration capacity. We introduce the use of a combination of parameters obtained from the cycler software PD Link with HomeChoicePro (Baxter International Inc., Illinois, United States) cyclers for predicting catheter dysfunction in automated peritoneal dialysis patients (APD). Out of 117 patients treated at the Medical University of Vienna between 2015 and 2021, we retrospectively identified all patients with verified catheter dysfunction (n = 14) and compared them to controls without clinical evidence of mechanical catheter problems and a recent X-ray confirming PD catheter tip in the rectovesical/rectouterine space (n = 19). All patients had a coiled single-cuff PD catheter, performed tidal PD, and received neutral pH bicarbonate/lactate-buffered PD fluids with low-glucose degradation products on APD. Icodextrin-containing PD fluids were used for daytime dwells. We retrieved cycler data for seven days each and tested parameters' predictive capability of catheter dysfunction. Total number of alarms/week > 7 as single predictive parameter of catheter dislocation identified 85.7% (sensitivity) of patients with dislocated catheter, whereas 31.6% (1-specificity) of control patients were false positive. A combination of parameters (number of alarms/week > 7, total drain time > 22 min, ultrafiltration of last fill < 150 mL) where at least two of three parameters appeared identified the same proportion of patients with catheter dislocation, but was more accurate in identifying controls (21.1% false positive). In contrast to yearly PET measurements, an easily applicable combination of daily cycler readout parameters, also available in new APD systems connected to remote monitoring platforms shows potential for diagnosis of catheter dysfunction during routine follow-up.

Funder

H2020 Marie Skłodowska-Curie Actions

Austrian Federal Ministry of Science, Research and Economy

National Foundation for Research, Technology and Development

Publisher

Springer Science and Business Media LLC

Subject

Multidisciplinary

Cited by 2 articles. 订阅此论文施引文献 订阅此论文施引文献,注册后可以免费订阅5篇论文的施引文献,订阅后可以查看论文全部施引文献

同舟云学术

1.学者识别学者识别

2.学术分析学术分析

3.人才评估人才评估

"同舟云学术"是以全球学者为主线,采集、加工和组织学术论文而形成的新型学术文献查询和分析系统,可以对全球学者进行文献检索和人才价值评估。用户可以通过关注某些学科领域的顶尖人物而持续追踪该领域的学科进展和研究前沿。经过近期的数据扩容,当前同舟云学术共收录了国内外主流学术期刊6万余种,收集的期刊论文及会议论文总量共计约1.5亿篇,并以每天添加12000余篇中外论文的速度递增。我们也可以为用户提供个性化、定制化的学者数据。欢迎来电咨询!咨询电话:010-8811{复制后删除}0370

www.globalauthorid.com

TOP

Copyright © 2019-2024 北京同舟云网络信息技术有限公司
京公网安备11010802033243号  京ICP备18003416号-3