Author:
Shin Young In,Kim Young Kook,Jeoung Jin Wook,Park Ki Ho
Abstract
AbstractThis study investigated intraocular pressure (IOP) in Dutch belted rabbits using two different tonometers, rebound tonometry (TonoVet Plus; TVP) and a Tonopen (Tono-Pen AVIA Vet; TPA). Post-pubescent male Dutch belted rabbits aged 36 weeks (n = 10 animals) were used in the study. IOP measurements were conducted every 2 weeks for 22 weeks using TVP and TPA on both eyes of each rabbit. The average IOP measurements were compared by the paired Student’s t-test. Pairwise Pearson’s correlation coefficients and Bland–Altman statistics were used. The overall mean IOP measured with TPA was significantly higher than that with TVP (23.5 ± 4.9 vs. 21.8 ± 2.4 mmHg for the right eyes; P = 0.045, and 23.0 ± 4.7 vs. 21.5 ± 2.4 mmHg for the left eyes; P = 0.047). Both tonometers tended to show increased IOP readings with age, and positive correlations between IOP and age were observed with both TPA (r = 0.95, P < 0.001 for right eyes; r = 0.95, P < 0.001 for left eyes) and TVP (r = 0.91, P < 0.001 for right eyes; r = 0.64, P = 0.024 for left eyes). The average bias calculated by subtracting TPA from TVP was − 1.60 (95% confidence intervals − 1.927, − 1.281) mmHg. IOP in post-pubescent Dutch belted rabbits tended to increase with age throughout the 22 week study.
Funder
Ministry of Trade, Industry and Energy
Korea Institute for Advancement of Technology
Publisher
Springer Science and Business Media LLC