Abstract
AbstractApelin exists in many isoforms, both in the circulation and in specific tissues. Apelin peptides have a short half-life but preservation before measurement is scarcely studied. Reproducible mass spectrometry methods to specifically measure a broad range of apelinergic peptide isoforms are currently lacking. A sample protocol to conserve apelinergic peptides in the preanalytical phase and a high-performance liquid chromatography–tandem mass spectrometry (HPLC–MS/MS) method to measure apelinergic isoforms was developed. Apelin was measured in plasma. For validation, human embryonic kidney (HEK) cells transfected with cDNA for preproapelin were used. Results were compared with a validated radioimmunoassay (RIA) method. Acidifying plasma to pH 2.5 improves post-sampling stability of apelin. HPLC–MS/MS was unable to detect apelin isoforms in plasma of healthy volunteers (n = 16) and chronic kidney disease patients (n = 4). RIA could detect apelin in concentrations between 71 and 263 fmol/l in 10 healthy volunteers. An optimized preanalytical protocol was developed. A sensitive and specific HPLC–MS/MS method failed to detect apelin in human plasma. Apelin-36 was detected in HEK cells transfected with cDNA for preproapelin. Currently, RIA with relatively selective antibodies is the best alternative for the measurement of apelin but novel sensitive and specific methods are needed.