Author:
Mouro João,Paoletti Paolo,Sartore Marco,Tiribilli Bruno
Abstract
AbstractWe investigate the response of a digitally controlled and parametrically pumped microcantilever used for sensing in a Phase-Locked Loop (PLL). We develop an analytical model for its dynamical response and obtain an explicit dependence on the rheological parameters of the surrounding viscous medium. Linearization of this model allows to find improved responsivity to density variations in the case of parametric suppression. Experiments with a commercial microcantilever validate the model, but also reveal an increase of frequency noise in the PLL associated with the parametric gain and phase, which, in most cases, restricts the attainable limit of detection. The noise in open-loop is studied by measuring the random fluctuations of the noise-driven deflection of the microcantilever, and a model for the power spectral density of amplitude, phase and frequency noises is discussed and used to explain the frequency fluctuations in the closed-loop PLL. This work concludes that parametric pumping in a PLL does not improve the sensing performance in applications requiring detecting frequency shifts.
Funder
Horizon 2020 Framework Programme
Publisher
Springer Science and Business Media LLC
Cited by
4 articles.
订阅此论文施引文献
订阅此论文施引文献,注册后可以免费订阅5篇论文的施引文献,订阅后可以查看论文全部施引文献