Abstract
AbstractPrevalent PET imaging reconstructs 2γ-photon pairs emitted after an annihilation from para-positronium (p-Ps) and rejects 3γ events from ortho-positronium (o-Ps) as noises. The 3γ/2γ decay ratio is ~ 3/7 in human body theoretically but in fact significantly lower due to pick-off process, hence PET imaging quality is well controlled. In a PET-MR hybrid unit, the MR magnetic field alters positronium decay patterns through magnetic quenching: all o-Ps and excited p-Ps states are split into finer quantum states under strong magnetic field, thus transitions between some triplet and singlet finer states (mz = 0) were no longer forbidden, thus some o-Ps converts to p-Ps spontaneously by emitting hyperfine split (HFS) photons, which also drops 3γ/2γ ratio hence helps PET imaging quality. However, inverse magnetic quenching might also occur if any external source of HFS frequencies is nearby, thus many p-Ps convert to o-Ps by absorbing those HFS photons (induced HFS transitions). This will dramatically increase 3γ/2γ ratio and hence degrade PET imaging quality instantaneously. The HFS spectrum lies in a broad range of microwaves, from 0.02 to 200 GHz. To prevent inverse magnetic quenching, it is necessary to block external microwave sources outside the hybrid vault, by adding a thin metal layer at all directions of the vault. This could be achieved by adopting the metallic Faraday Cage, which was originally for MR shielding, with possible amendment if necessary. The frequencies of excitation pulses in MR imaging overlap with HFS spectrum, however, the chance for mutual interference during hybrid imaging is small, hence there seems no need to veto each other during hybrid scans.