The comparison of the process of manual and robotic positioning of the electrode performing radiofrequency ablation under the control of a surgical navigation system

Author:

Levin A. A.,Klimov D. D.,Nechunaev A. A.,Vorotnikov A. A.,Prokhorenko L. S.,Grigorieva E. V.,Astakhov D. A.,Poduraev Y. V.,Panchenkov D. N.

Abstract

AbstractThis study is aimed at the comparison of the process of manual and robotic positioning of the electrode performing radiofrequency ablation under the control of a surgical navigation system. The main hypothesis of this experiment was that the use of a collaborative manipulator (KUKA iiwa) will allow to position the active part of the electrode relative to the center of the tumor more accurately and from the first attempt. We also monitor the stability of the electrode′s velocity during insertion and consider some advantages in ergonomics using the robotic manipulator. We use three more criteria to compare the surgeon's and robotic performance, unlike other studies, where only the target point's accuracy criterion is observed. The main idea is to examine the movement parameters of the electrode that can lead to potential patient trauma. Sphere-shaped tumor phantoms measuring 8 mm in diameter were filled with contrast and inserted in bovine livers. 10 livers were used for the robotic experiment and an equal quantity for manual surgery. The livers were encased in silicone phantoms designed to imitate the liver position in a real patient's abdominal cavity. Analysis of CT data gave the opportunity to find the entry and the target point for each tumor phantom. This data was loaded into a surgical navigation system that was used to track and record the position of the RF-electrode during the operation for further analysis. The standard deviation of points from the programmed linear trajectory totaled in the average 0.3 mm for the robotic experiment and 2.33 mm for the manual operation with a maximum deviation of 0.55 mm and 7.99 mm respectively. Standard deviation from the target point was 2.69 mm for the collaborative method and 2.49 mm for the manual method. The average velocity was 2.97 mm/s for the manipulator and 3.12 mm/s for the manual method, but the standard deviation of the velocity relative to the value of the average velocity was 0.66 mm/s and 3.05 mm/s respectively. Thus, in two criteria out of three, the manipulator is superior to the surgeon, and equality is established in one. Surgeons also noticed advantages in ergonomics performing the procedure using the manipulator. This experiment was produced as part of the work on the developing of a robotic multifunctional surgical complex. We can confirm the potential advantages of using collaborative robotic manipulators for minimally invasive surgery in case of practice for cancer treatment.

Publisher

Springer Science and Business Media LLC

Subject

Multidisciplinary

Cited by 9 articles. 订阅此论文施引文献 订阅此论文施引文献,注册后可以免费订阅5篇论文的施引文献,订阅后可以查看论文全部施引文献

同舟云学术

1.学者识别学者识别

2.学术分析学术分析

3.人才评估人才评估

"同舟云学术"是以全球学者为主线,采集、加工和组织学术论文而形成的新型学术文献查询和分析系统,可以对全球学者进行文献检索和人才价值评估。用户可以通过关注某些学科领域的顶尖人物而持续追踪该领域的学科进展和研究前沿。经过近期的数据扩容,当前同舟云学术共收录了国内外主流学术期刊6万余种,收集的期刊论文及会议论文总量共计约1.5亿篇,并以每天添加12000余篇中外论文的速度递增。我们也可以为用户提供个性化、定制化的学者数据。欢迎来电咨询!咨询电话:010-8811{复制后删除}0370

www.globalauthorid.com

TOP

Copyright © 2019-2024 北京同舟云网络信息技术有限公司
京公网安备11010802033243号  京ICP备18003416号-3