AbstractAccurately monitoring the extent of freezing in biological tissue is an important requirement for cryoablation, a minimally invasive cancer treatment that induces cell death by freezing tissue with a cryoprobe. During the procedure, monitoring is required to avoid unnecessary harm to the surrounding healthy tissue and to ensure the tumor is properly encapsulated. One commonly used monitoring method is attenuation-based computed tomography (CT), which visualizes the ice ball by utilizing its hypoattenuating properties compared to unfrozen tissue. However, the contrast between frozen and unfrozen tissue remains low. In a proof-of-principle experiment, we show that the contrast between frozen and unfrozen parts of a porcine phantom mimicking breast tissue can be greatly enhanced by acquiring X-ray dark-field images that capture the increasing small-angle scattering caused by the ice crystals formed during the procedure. Our results show that, compared to X-ray attenuation, the frozen region is detected significantly better in dark-field radiographs and CT scans of the phantom. These findings demonstrate that X-ray dark-field imaging could be a potential candidate for improved monitoring of cryoablation procedures.
Funder
Deutsche Forschungsgemeinschaft (DFG), Research Training Group
Technische Universität München
Publisher
Springer Science and Business Media LLC
Reference45 articles.
1. Mahnken, A. H., König, A. M. & Figiel, J. H. Current technique and application of percutaneous cryotherapy. In RöFo-Fortschritte auf dem Gebiet der Röntgenstrahlen und der bildgebenden Verfahren Vol. 190 (eds Mahnken, A. H. et al.) 836–846 (Georg Thieme Verlag KG, 2018).
2. Pusceddu, C., Paliogiannis, P., Nigri, G. & Fancellu, A. Cryoablation in the management of breast cancer: Evidence to date. Breast Cancer Targets Ther. 1, 283–292 (2019).
3. Rewcastle, J. C., Sandison, G. A., Muldrew, K., Saliken, J. C. & Donnelly, B. J. A model for the time dependent three-dimensional thermal distribution within iceballs surrounding multiple cryoprobes. Med. Phys. 28, 1125–1137 (2001).
4. Niu, L.-Z., Li, J.-L. & Xu, K.-C. Percutaneous cryoablation for liver cancer. J. Clin. Transl. Hepatol. 2, 182 (2014).
5. Poplack, S. P. et al. A pilot study of ultrasound-guided cryoablation of invasive ductal carcinomas up to 15 mm with mri follow-up and subsequent surgical resection. Am. J. Roentgenol. 204, 1100 (2015).