Author:
Miyazu Takahiro,Osawa Satoshi,Tamura Satoshi,Tani Shinya,Ishida Natsuki,Matsuura Tomoharu,Yamade Mihoko,Iwaizumi Moriya,Hamaya Yasushi,Furuta Takahisa,Sugimoto Ken
Abstract
AbstractIn 2012, Japan approved the use of a tag-less patency capsule (PC), which evaluates gastrointestinal patency before small-bowel capsule endoscopy (SBCE). This study aimed to evaluate the validity of our modification on the passage criteria for this PC in clinical practice. We retrospectively enrolled 326 consecutive patients who underwent PC examination before SBCE. If X-ray could not reveal the PC in the body during the judgement time (30–33 h after ingestion), we defined it as ‘estimated patency’ and performed SBCE. We employed plain computed tomography (CT) for the second judgement, as needed. The overall patency rate was 95.1%. By X-ray, 41 (12.6%) patients were judged to have ‘estimated patency’, and SBCE could be safely performed. Plain CT judgement was necessary in 106 patients (32.5%). One PC case had a residual coating film associated with stenosis in a patient with Crohn’s disease (CD), and one (0.3%) SBCE case had capsule retention resulting from false CT judgement. Multivariate analysis revealed that established CD and inpatient were factors related to no-patency. In conclusion, PC is useful for examining gastrointestinal patency, keeping in mind CT misjudgement. If PC was not found in the body via X-ray, performing SBCE as ‘estimated patency’ seemed appropriate.
Publisher
Springer Science and Business Media LLC
Cited by
5 articles.
订阅此论文施引文献
订阅此论文施引文献,注册后可以免费订阅5篇论文的施引文献,订阅后可以查看论文全部施引文献