Author:
Yamaguchi Yukiko,Uesato Masaya,Yonemoto Shohei,Maruyama Tetsuro,Urahama Ryuma,Suito Hiroshi,Kishimoto Takashi,Shiko Yuki,Ozawa Yoshihito,Kawasaki Yohei,Matsubara Hisahiro
Abstract
AbstractOne of the complications of esophageal endoscopic submucosal dissection (ESD) is postoperative stricture formation. Stenosis formation is associated with inflammation and fibrosis in the healing process. We hypothesized that the degree of thermal damage caused by the device is related to stricture formation. We aimed to reveal the relationship between thermal damage and setting value of the device. We energized a resected porcine esophagus using the ESD device (Flush Knife 1.5). We performed 10 energization points for 1 s, 3 s, and 5 s at four setting values of the device. We measured the amount of current flowing to the conducted points and the temperature and evaluated the effects of thermal damage pathologically. As results, the mean highest temperatures for 1 s were I (SWIFT Effect3 Wat20): 61.19 °C, II (SWIFT Effect3 Wat30): 77.28 °C, III (SWIFT Effect4 Wat20): 94.50 °C, and IV (SWIFT Effect4 Wat30): 94.29 °C. The mean heat denaturation areas were I: 0.84 mm2, II: 1.00 mm2, III: 1.91 mm2, and IV: 1.54 mm2. The mean highest temperature and mean heat denaturation area were significantly correlated (P < 0.001). In conclusion, Low-current ESD can suppress the actual temperature and thermal damage in the ESD wound.
Publisher
Springer Science and Business Media LLC
Cited by
1 articles.
订阅此论文施引文献
订阅此论文施引文献,注册后可以免费订阅5篇论文的施引文献,订阅后可以查看论文全部施引文献