Author:
Tsuchida Wakako,Yamakoshi Yoshiki,Matsuo Shingo,Asakawa Mayu,Sugahara Keita,Fukaya Taizan,Yamanaka Eiji,Asai Yuji,Nitta Naotaka,Ooie Toshihiko,Suzuki Shigeyuki
Abstract
AbstractIn recent years, non-invasive measurement of tissue stiffness (hardness) using ultrasound elastography has attracted considerable attention. It has been used to evaluate muscle stiffness in the fields of rehabilitation, sports, and orthopedics. However, ultrasonic diagnostic devices with elastography systems are expensive and clinical use of such devices has been limited. In this study, we proposed a novel estimation method for vibration-based shear wave elastography measurement of human skeletal muscle, then determined its reproducibility and reliability. The coefficient of variation and correlation coefficient were used to determine reproducibility and reliability of the method by measuring the shear wave velocities in konjac phantom gels and agar phantom gels, as well as skeletal muscle. The intra-day, day-to-day, and inter-operator reliabilities were good when measuring the shear wave velocities in phantom gels. The intra-day and day-to-day reliabilities were good when measuring the shear wave velocities in skeletal muscle. The findings confirmed adequate reproducibility and reliability of the novel estimation method for vibration-based shear wave elastography. Therefore, the proposed measurement method may be a useful tool for evaluation of muscle stiffness.
Funder
Japan Society for the Promotion of Science
Publisher
Springer Science and Business Media LLC
Cited by
10 articles.
订阅此论文施引文献
订阅此论文施引文献,注册后可以免费订阅5篇论文的施引文献,订阅后可以查看论文全部施引文献