AbstractDry electroencephalography (EEG) electrodes provide rapid, gel-free, and easy EEG preparation, but with limited wearing comfort. We propose a novel dry electrode comprising multiple tilted pins in a flower-like arrangement. The novel Flower electrode increases wearing comfort and contact area while maintaining ease of use. In a study with 20 volunteers, we compare the performance of a novel 64-channel dry Flower electrode cap to a commercial dry Multipin electrode cap in sitting and supine positions. The wearing comfort of the Flower cap was rated as significantly improved both in sitting and supine positions. The channel reliability and average impedances of both electrode systems were comparable. Averaged VEP components showed no considerable differences in global field power amplitude and latency, as well as in signal-to-noise ratio and topography. No considerable differences were found in the power spectral density of the resting state EEGs between 1 and 40 Hz. Overall, our findings provide evidence for equivalent channel reliability and signal characteristics of the compared cap systems in the sitting and supine positions. The reliability, signal quality, and significantly improved wearing comfort of the Flower electrode allow new fields of applications for dry EEG in long-term monitoring, sensitive populations, and recording in supine position.