Abstract
AbstractTube-shaped blood vessel models that mimic their geometries and mechanical properties can deliver reliable and realistic behavioral information such as deformation and rupture during procedures such as insertion of medical devices. Thickness of vessel walls is an important parameter for fabricating the blood vessel models owing to their strong influence on the model behavior, especially during deformation. The dip-coating method is used to fabricate blood vessel models; however, non-uniform wall thicknesses are observed using this method. This study aimed at finding the characteristics of stereo “angular control dip-coating” (ACDC) system to develop a dip-coating system that can produce tubular models with uniformed wall thickness. The system developed here enables an observation of the substrate behavior from two different views. The conditions of dip-coating used in this study produce 1.36–1.82 mm in the maximum and 0.188–0.435 mm in minimum wall thickness and the fabricated walls cover the realistic range of carotid arterial dimensions. The characteristics of the ACDC system indicate that ACDC is effective for fabricating the uniform wall thickness particularly in the strong curved parts.
Funder
ImPACT Program of Council for Science, Technology and Innovation (Cabinet Office, Government of Japan).
Publisher
Springer Science and Business Media LLC
Cited by
8 articles.
订阅此论文施引文献
订阅此论文施引文献,注册后可以免费订阅5篇论文的施引文献,订阅后可以查看论文全部施引文献