AbstractNeurofibromatosis type 2 is a neurogenetic disorder with an incidence of about 1:33.000. Hallmarks are bilateral benign vestibular schwannomas, which can lead to deafness or brainstem compression. Volumetric tumor measurements are essential to assess the efficacy of new therapies. We present a statistical and methodical comparison of three volumetric image analysis tools. We performed volumetric measurements on phantoms with predefined volumes (0.1 to 8.0 ml) and tumors seen on 32 head MRI scans from eight NF2 patients with BrainLab, ITK-Snap, or OsiriX. The software was compared with regard to accuracy and reproducibility of the measurements and time required for analysis. The mean volume estimated by all three software programs differed significantly from the true volume of the phantoms, but OsiriX and BrainLab gave estimates that were not significantly different from each other. For the actual tumors, the estimated volumes with all three software tools showed a low coefficient of variability, but the mean volume estimates differed among the tools. OsiriX showed the shortest analysis time. Volumetric assessment of MRI images is associated to an intrinsic risk of miscalculation. For precise volumes it is mandatory to use the same volumetric tools for all measurements.