Author:
Wakabayashi Ryo,Shiko Yuki,Kodaira Tomofumi,Shiroshita Yuko,Otsuka Hitomi,Baba Kosuke,Hishinuma Norimasa
Abstract
AbstractThe aim of this randomized controlled trial was to determine the efficacy of stylet angulation at the holding position during tracheal intubation with a McGRATH MAC videolaryngoscope. Patients were randomized to a group for intubation without stylet angulation at the holding position (non-angulation group) and to a group for intubation with stylet angulation at the holding position (angulation group). The primary outcome was the time for placement of the tracheal tube. Sixty patients were analyzed. The mean (standard deviation) times for tube placement were 21.3 (5.6) s in the non-angulation group and 16.9 (3.8) s in the angulation group (P < 0.001). The scores of operator's perception of difficulty in tube delivery, number of attempts for tube delivery, and degrees of extension, abduction, internal rotation of the right upper arm and extension of the right wrist during tube placement in the angulation group were significantly smaller than those in the non-angulation group (P < 0.001, P = 0.002, P < 0.001, P < 0.001, P < 0.001, P < 0.001, respectively). Our results suggest that stylet angulation at the holding position improves maneuverability of the tracheal tube and enables easy, smooth, and swift tube placement during tracheal intubation with a McGRATH MAC videolaryngoscope.