Author:
Stino Heiko,Riessland Susanna,Sedova Aleksandra,Datlinger Felix,Sacu Stefan,Schmidt-Erfurth Ursula,Pollreisz Andreas
Abstract
AbstractComparison of two ultra-widefield (UWF) color-fundus (CF) imaging devices in diabetic patients for visualization of retinal periphery and detection of early microvascular lesions. The total gradable areas (TGA) seen on non-mydriatic CF-images of two UWF-imaging devices (Optos Daytona P200T; Zeiss Clarus 700) were compared and differences in projected area measured. Retinal periphery outside the 7 standard fields (7SF) was divided into: F3 temporal, F4 superotemporal, F5 inferotemporal, F6 superonasal, F7 inferonasal. DR stage was evaluated in the 7SF and the TGA on images of both devices and compared using Cohens κ. 67 eyes of 67 patients (52.5 ± 15.3 years) were analysed. DR stages in the 7SF were no (n = 36 Optos, n = 35 Clarus), mild (n = 16 Optos, n = 17 Clarus), and moderate DR (n = 15). Optos depicted significantly more area in F3 (median [interquartile range]; 2.41% [1.06–4.11] vs 0% [0–0], P < 0.001) and Clarus in F7 (3.29% [0–7.69] vs 0% [0–3.27], P = 0.002). In 4 eyes DR-stage was higher using Optos due to peripheral lesions not seen on the Clarus. Interrater reliability of DR-stage on both devices was almost perfect in the 7SF (κ = 0.975) and the TGA (κ = 0.855). Reliability in detecting signs of early DR is high on both devices. Clarus allowed for better visualization of the inferonasal field, Optos of the temporal field.
Publisher
Springer Science and Business Media LLC
Cited by
6 articles.
订阅此论文施引文献