Abstract
Abstract
Purpose
To compare 10-MHz and 15-MHz B-scan probes regarding the detection and localization of retinal detachment (RD) in silicone-oil-filled eyes.
Methods
This cross-sectional observational study included 100 eyes (98 patients) -having media opacity precluding fundus examination- scheduled for silicone-oil removal. Patients were examined in the sitting-position using both frequencies one week preoperatively. Longitudinal and transverse scans were taken in primary-gaze, inferior, inferonasal, and inferotemporal positions to detect the presence/absence and extent of RD. Patients were sub-grouped according to axial lengths (AXLs), state of silicone emulsification, and globe filling. Agreement between sonographic and intraoperative observations was compared.
Results
No statistically significant differences were found between 15-MHz and intra-operative findings regarding RD detection (P = 0.752) and accurate localization of inferior, inferonasal, and inferotemporal RD (P = 0.279, 0.606, and 0.599). There were statistically significant differences between 10-MHz and intra-operative findings regarding RD detection and localization (P < 0.001). The 15-MHz probe was superior to 10-MHz probe regarding the accuracy of RD detection and localization (94% and 47%, respectively). The accuracy of 15-MHz probe was 88%, 83%, and 85% in detecting and localizing inferior, inferonasal, and inferotemporal RD compared to 45%, 60%, and 62% with 10-MHz probe. The 15-MHz probe showed better sensitivity while 10-MHz probe showed better accuracy in eyes with short AXLs. The 10-MHz probe showed better sensitivity in patients with sonographic emulsification while15-MHz probe had better sensitivity in detecting vitreoretinal-interface disorders.
Conclusion
The 15-MHz B-scan probe is more accurate in detecting and localizing recurrent RD in silicone-oil-filled globes with higher sensitivity in detecting vitreoretinal-interface disorders.
Publisher
Springer Science and Business Media LLC
Subject
General Arts and Humanities,Ophthalmology,Sensory Systems
Reference15 articles.
1. Mundt GH Jr, Hughes WF Jr. Ultrasonics in ocular diagnosis. Am J Ophthalmol. 1956;41:488–98.
2. Li YF, Li DJ, Wang ZY, Chen W, Zhao Q, Cui R, et al. Ultrasonic diagnosis of retinal detachment in eyes with silicone oil tamponade. Chin J Ophthalmol. 2017;53:842–6.
3. Kumar A, Sharma N, Singh R. Prone position ultrasonography in silicone-filled eyes. Acta Ophthalmol Scand. 1998;76:496–8.
4. Hewick SA, Fairhead AC, Culy JC, Atta HR. A comparison of 10 MHz and 20 MHz ultrasound probes in imaging the eye and orbit. Br J Ophthalmol. 2004;88:551–5.
5. Agarwal A, Maleki A, Nguyen QD. History and principles of ocular ultrasonography. In Clinical Atlas of Ophthalmic Ultrasound 2019 (pp. 1–6). Springer, Cham.