Author:
O’Reilly Megan,Lee Tze-Fun,Cheung Po-Yin,Schmölzer Georg M.
Abstract
Abstract
Background
High-quality chest compressions (CC) are an important factor of neonatal resuscitation. Mechanical CC devices may provide superior CC delivery and improve resuscitation outcomes. We aimed to compare the hemodynamic effects of CC delivered by machine and human using a neonatal piglet model.
Methods
Twelve asphyxiated piglets were randomized to receive CC during resuscitation using an automated mechanical CC device (“machine”) or the two-thumb encircling technique (“human”). CC was superimposed with sustained inflations.
Results
Twelve newborn piglets (age 0–3 days, weight 2.12 ± 0.17 kg) were included in the study. Machine-delivered CC resulted in an increase in stroke volume, and minimum and maximum rate of left ventricle pressure change (dp/dtmin and dp/dtmax) compared to human-delivered CC.
Conclusions
During machine-delivered CC, stroke volume and left ventricular contractility were significantly improved. Mechanical CC devices may provide improved cardiopulmonary resuscitation outcomes in neonatal cardiac arrest induced by asphyxia.
Impact
Machine chest compression leads to changes in hemodynamic parameters during resuscitation of asphyxiated neonatal piglets, namely greater stroke volume and left ventricular contractility, compared with standard two-thumb compression technique. Mechanical chest compression devices may provide improved cardiopulmonary resuscitation outcomes in neonatal and pediatric asphyxia-induced cardiac arrest.
Publisher
Springer Science and Business Media LLC
Subject
Pediatrics, Perinatology and Child Health
Cited by
2 articles.
订阅此论文施引文献
订阅此论文施引文献,注册后可以免费订阅5篇论文的施引文献,订阅后可以查看论文全部施引文献