Titus Jitto,Wu Alan H. B.,Biswal Siddharth,Burman Atandra,Sengupta Shantanu P.,Sengupta Partho P.
Abstract
Abstract
Background
The levels of circulating troponin are principally required in addition to electrocardiograms for the effective diagnosis of acute coronary syndrome. Current standard-of-care troponin assays provide a snapshot or momentary view of the levels due to the requirement of a blood draw. This modality further restricts the number of measurements given the clinical context of the patient. In this communication, we present the development and early validation of non-invasive transdermal monitoring of cardiac troponin-I to detect its elevated state.
Methods
Our device relies on infrared spectroscopic detection of troponin-I through the dermis and is tested in stepwise laboratory, benchtop, and clinical studies. Patients were recruited with suspected acute coronary syndrome.
Results
We demonstrate a significant correlation (r = 0.7774, P < 0.001, n = 52 biologically independent samples) between optically-derived data and blood-based immunoassay measurements with and an area under receiver operator characteristics of 0.895, sensitivity of 96.3%, and specificity of 60% for predicting a clinically meaningful threshold for defining elevated Troponin I.
Conclusion
This preliminary work introduces the potential of a bloodless transdermal measurement of troponin-I based on molecular spectroscopy. Further, potential pitfalls associated with infrared spectroscopic mode of inquiry are outlined including requisite steps needed for improving the precision and overall diagnostic value of the device in future studies.
Publisher
Springer Science and Business Media LLC
Cited by
14 articles.
订阅此论文施引文献
订阅此论文施引文献,注册后可以免费订阅5篇论文的施引文献,订阅后可以查看论文全部施引文献