Evaluation of the ShuntCheck Noninvasive Thermal Technique for Shunt Flow Detection in Hydrocephalic Patients

Author:

Madsen Joseph R.1,Abazi Gani S.1,Fleming Laurel.1,Proctor Mark.1,Grondin Ron.12,Magge Suresh.13,Casey Peter.1,Anor Tomer.1

Affiliation:

1. Department of Neurosurgery, Children's Hospital Boston, Harvard Medical School, Boston, Massachusetts

2. Current address: Department of Neurosurgery, Nationwide Children's Hospital, Columbus, Ohio

3. Current address: Division of Neurosurgery, Children's National Medical Center, Washington, District of Columbia

Abstract

Abstract BACKGROUND: ShuntCheck (Neuro Diagnostic Devices, Inc., Trevose, Pennsylvania) is a new device designed to detect cerebrospinal fluid (CSF) flow in a shunt by sensing skin temperature downstream from a region of CSF cooled by an ice cube. OBJECTIVE: To understand its accuracy and utility, we evaluated the use of this device during routine office visits as well as during workup for suspected shunt malfunction. METHODS: One hundred shunted patients were tested, including 48 evaluated during possible shunt malfunction, of whom 24 went on to surgical exploration. Digitally recorded data were blindly analyzed and compared with surgical findings and clinical follow-up. RESULTS: Findings in the 20 malfunctioning shunts with unambiguous flow or absence of flow at surgery were strongly correlated with ShuntCheck results (sensitivity and specificity to flow of 80% and 100%, respectively, P = .0007, Fisher's exact test, measure of agreement κ = 0.8). However, the thermal determination did not distinguish patients in the suspected malfunction group who received surgery from those who were discharged without surgery (P = .248 by Fisher's exact test, κ = 0.20). Half of the patients seen in routine office visits did not have detectable flow, although none required shunt revision on clinical grounds. Intermittent flow was specifically demonstrated in one subject who had multiple flow determinations. CONCLUSION: Operative findings show that the technique is sensitive and specific for detecting flow, but failure to detect flow does not statistically predict the need for surgery. A better understanding of the normal dynamics of flow in individual patients, which this device may yield, will be necessary before the true clinical utility of non-invasive flow measurement can be assessed.

Publisher

Ovid Technologies (Wolters Kluwer Health)

Subject

Neurology (clinical),Surgery

Reference35 articles.

1. Time trends and demographics of deaths from congenital hydrocephalus in children in the United States: National Center for Health Statistics data, 1979 to 1998;Chi;J Neurosurg,2005

2. Dynamic measurement of the flow rate in cerebrospinal fluid shunts in hydrocephalic patients;Hidaka;Eur J Nucl Med,2001

3. What we don't (but should) know about hydrocephalus;Bergsneider;J Neurosurg. Mar,2006

4. Cerebrospinal fluid pulsatility and hydrocephalus: the fourth circulation;Madsen;Clin Neurosurg,2006

5. Shunt technology: contemporary concepts and prospects;Scott;Clin Neurosurg,2003

Cited by 31 articles. 订阅此论文施引文献 订阅此论文施引文献,注册后可以免费订阅5篇论文的施引文献,订阅后可以查看论文全部施引文献