A Study of the Accuracy of CyberKnife Spinal Radiosurgery Using Skeletal Structure Tracking

Author:

Ho Anthony K.1,Fu Dongshan2,Cotrutz Cristian1,Hancock Steven L.1,Chang Steven D.3,Gibbs Iris C.1,Maurer Calvin R.2,Adler John R.3

Affiliation:

1. Department of Radiation Oncology, Stanford University Medical Center, Stanford, California

2. Accuray, Inc., Sunnyvale, California

3. Department of Neurosurgery, Stanford University Medical Center, Stanford, California

Abstract

Abstract Objective: New technology has enabled the increasing use of radiosurgery to ablate spinal lesions. The first generation of the CyberKnife (Accuray, Inc., Sunnyvale, CA) image-guided radiosurgery system required implanted radiopaque markers (fiducials) to localize spinal targets. A recently developed and now commercially available spine tracking technology called Xsight (Accuray, Inc.) tracks skeletal structures and eliminates the need for implanted fiducials. The Xsight system localizes spinal targets by direct reference to the adjacent vertebral elements. This study sought to measure the accuracy of Xsight spine tracking and provide a qualitative assessment of overall system performance. Methods: Total system error, which is defined as the distance between the centroids of the planned and delivered dose distributions and represents all possible treatment planning and delivery errors, was measured using a realistic, anthropomorphic head-and-neck phantom. The Xsight tracking system error component of total system error was also computed by retrospectively analyzing image data obtained from eleven patients with a total of 44 implanted fiducials who underwent CyberKnife spinal radiosurgery. Results: The total system error of the Xsight targeting technology was measured to be 0.61 mm. The tracking system error component was found to be 0.49 mm. Conclusion: The Xsight spine tracking system is practically important because it is accurate and eliminates the use of implanted fiducials. Experience has shown this technology to be robust under a wide range of clinical circumstances.

Publisher

Ovid Technologies (Wolters Kluwer Health)

Subject

Clinical Neurology,Surgery

Reference52 articles.

1. Surgery for tic douloureaux;Apfelbaum;Clin Neurosurg,1983

2. Endoscopically assisted minimally invasive microvascular decompression of hemifacial spasm;Badr-El-Dine;Otol Neurotol,2002

3. Trigeminal neuralgia: Results of microsurgical and endoscopic-assisted vascular decompression [in French];Balansard ChF, Meller;Ann Otolaryngol Chir Cervicofac,2003

4. The long-term outcome of microvascular decompression for trigeminal neuralgia;Barker FG 2nd, Jannetta;N Engl J Med,1996

5. Endoscopic-assisted ear surgery;Bottrill;Am J Otol,1995