Platinum-Iridium Subdermal Magnetic Resonance Imaging-Compatible Needle Electrodes Are Suitable for Intraoperative Neurophysiological Monitoring During Image-Guided Surgery With High-Field Intraoperative Magnetic Resonance Imaging: An Experimental Study

Author:

D’Andrea Giancarlo1,Angelini Albina1,Foresti Camillo2,Familiari Pietro1,Caroli Emanuela1,Frati Alessandro3

Affiliation:

1. Neurosurgery, Department of NESMOS, Faculty of Medicine and Psychology

2. Department of Neurology, Unit of Neurophysiopathology, Papa Giovanni XXIII Hospital, Bergamo, Italy

3. Department of Neurosurgery, IRCCS Neuromed Pozzilli (Is), University of Rome—Sapienza, Rome, Italy

Abstract

Abstract BACKGROUND: Neurosurgery aims to achieve maximal tumor resection while preserving neurological function. Tools such as neuronavigation, high-field intraoperative magnetic resonance imaging (iMRI), and intraoperative neurophysiological monitoring (IOM) have consistently helped to achieve this goal, but integration has often been difficult. Surgery of eloquent areas requires IOM, which in an operating theater equipped with high-field (1.5-T) iMRI could present several issues. OBJECTIVE: To identify the electrodes types more suitable for IOM in a high-field iMRI operating theater by performing an experimental study on phantoms, to report our experience with platinum-iridium (Pt/Ir) electrodes during surgery, and to prove that integration between IOM with Pt/Ir electrodes and high-field iMRI is safe and reliable. METHODS: Electrodes of different materials (gold, Pt/Ir, and stainless steel) were tested on jelly phantom and apples to evaluate their safety and compatibility. Subsequently, electrodes were tested on 5 healthy volunteers before being used on patients. RESULTS: None of the different electrodes presented thermal instability, and no damage to the volunteers’ skin occurred. Stainless steel electrodes caused severe imaging distortion. Gold electrodes had no distortion, but their high cost makes their use in routine surgery unaffordable. Pt/Ir electrodes are significantly less expensive than gold electrodes and were completely safe, compatible, and suitable for use in an operating theater with high-field iMRI, providing excellent IOM and mild interference that did not affect the quality of intraoperative imaging. CONCLUSION: We suggest the use of Pt/Ir electrodes for IOM in 1.5-T iMRI suites.

Publisher

Ovid Technologies (Wolters Kluwer Health)

Subject

Clinical Neurology,Surgery

Reference8 articles.

1. Intraoperative neurophysiological monitoring in an open low-field magnetic resonance imaging system: clinical experience and technical considerations;Szelènyi;Neurosurgery,2008

2. Utilization of intraoperative motor mapping in glioma surgery with high-field intraoperative magnetic resonance imaging;Hatiboglu;Stereotact Funct Neurosurg,2010

3. Cortical mapping and frameless stereotactic navigation in the high-field intraoperative magnetic resonance imaging suite;Weingarten;J Neurosurg,2009

4. Metallic neurosurgical implants: evaluation of magnetic field interactions, heating, and artifacts at 1.5-Tesla;Shellock;J Magn Reson Imaging,2001

5. Effect of 7.0 Tesla MRI on upper eyelid implants;Schrom;Ophthal Plast Reconstr Surg,2006