Abstract

BACKGROUND AND OBJECTIVES: We developed, tested, and validated a novel, noninvasive, Leksell G frame–based fiducial attachment, for use in stereotactic registration for stereoelectroencephalography (sEEG). Use of the device increased the number of fixed reference points available for registration, while obviating the need for additional scalp incisions. We report here on our experience and safety profile of using the device. METHODS: We collected registration data using the fiducial device across 25 adult and pediatric patients with epilepsy consecutively undergoing robotic-guided sEEG for invasive epilepsy monitoring, treated between May 2022 and July 2023. ROSA One Brain was used for trajectory planning and electrode implantation. Postoperative clinical and radiographic data were computed and quantified, including mean registration error for all patients. Entry point, target point (TP), and angular errors were measured. Descriptive statistics and correlation coefficients for error were calculated. RESULTS: Twenty-five patients underwent robotic-guided sEEG implantation (11 patients, bilateral; 10 patients, left unilateral; 4 patients, right). The mean number of electrodes per patient was 18 ± 3. The average mean registration error was 0.77 ± 0.11 mm. All patients were implanted with Ad-Tech depth electrodes. No clinically relevant complications were reported. Analysis of trajectory error was performed on 446 electrodes. The median entry point error was 1.03 mm (IQR 0.69-1.54). The median TP error was 2.26 mm (IQR 1.63-2.93). The mean angular error was 0.03 radians (IQR 0.02-0.05). There was no significant correlation between root mean square error and lead error. Root mean square error did not appreciably change over time, nor were there any significant changes in average angular, entry point, or TP error metrics. CONCLUSION: A novel, noninvasive, Leksell G frame–based fiducial attachment was developed, tested, and validated, facilitating O-arm–based stereotactic registration for sEEG. This simple innovation maintained an excellent accuracy and safety profile for sEEG procedures in epilepsy patients, with the added advantages of providing additional reference points for stereotactic registration, without requiring additional scalp incisions.

Publisher

Ovid Technologies (Wolters Kluwer Health)

同舟云学术

1.学者识别学者识别

2.学术分析学术分析

3.人才评估人才评估

"同舟云学术"是以全球学者为主线,采集、加工和组织学术论文而形成的新型学术文献查询和分析系统,可以对全球学者进行文献检索和人才价值评估。用户可以通过关注某些学科领域的顶尖人物而持续追踪该领域的学科进展和研究前沿。经过近期的数据扩容,当前同舟云学术共收录了国内外主流学术期刊6万余种,收集的期刊论文及会议论文总量共计约1.5亿篇,并以每天添加12000余篇中外论文的速度递增。我们也可以为用户提供个性化、定制化的学者数据。欢迎来电咨询!咨询电话:010-8811{复制后删除}0370

www.globalauthorid.com

TOP

Copyright © 2019-2024 北京同舟云网络信息技术有限公司
京公网安备11010802033243号  京ICP备18003416号-3