Affiliation:
1. Department of Neurosurgery, Osaka University Graduate School of Medicine, Suita, Japan;
2. Epilepsy Center, Osaka University Hospital, Suita, Japan
Abstract
BACKGROUND AND OBJECTIVES:
Boltless implantation of stereo-electroencephalography electrode is a useful alternative especially when anchor bolt is not available such as in country with limited resources or is less appropriate such as placement in patients with thin skull or at the occiput area, despite some drawbacks including potential dislodgement. While the accuracy of implantation using anchor bolt is well-studied, data on boltless implantation remain scarce. This study aimed to reveal the accuracy, permissible error for actual placement of electrodes within the grey matter, and delayed electrode dislodgement in boltless implantation.
METHODS:
A total of 120 electrodes were implanted in 15 patients using a Leksell Stereotactic G Frame with each electrode fixed on the scalp using sutures. Target point error was defined as the Euclidean distance between the planned target and the electrode tip on immediate postimplantation computed tomography. Similarly, delayed dislodgement was defined as the Euclidean distance between the electrode tips on immediate postimplantation computed tomography and delayed MRI. The factors affecting accuracy were evaluated using multiple linear regression. The permissible error was defined as the largest target point error that allows the maximum number of planned gray matter electrode contacts to be actually placed within the gray matter as intended.
RESULTS:
The median (IQR) target point error was 2.6 (1.7-3.5) mm, and the permissible error was 3.2 mm. The delayed dislodgement, with a median (IQR) of 2.2 (1.4-3.3) mm, was dependent on temporal muscle penetration (P = 5.0 × 10−4), scalp thickness (P < 5.1 × 10−3), and insertion angle (P = 3.4 × 10−3).
CONCLUSION:
Boltless implantation of stereo-electroencephalography electrode offers an accuracy comparable to those using anchor bolt. During the planning of boltless implantation, target points should be placed within 3.2 mm from the gray-white matter junction and a possible delayed dislodgement of 2.2 mm should be considered.
Funder
Ministry of Education, Culture, Sports, Science and Technology
Publisher
Ovid Technologies (Wolters Kluwer Health)