Artificial vision with wirelessly powered subretinal electronic implant alpha-IMS

Author:

Stingl Katarina1,Bartz-Schmidt Karl Ulrich1,Besch Dorothea1,Braun Angelika2,Bruckmann Anna1,Gekeler Florian1,Greppmaier Udo2,Hipp Stephanie1,Hörtdörfer Gernot3,Kernstock Christoph1,Koitschev Assen4,Kusnyerik Akos15,Sachs Helmut6,Schatz Andreas1,Stingl Krunoslav T.7,Peters Tobias8,Wilhelm Barbara8,Zrenner Eberhart1

Affiliation:

1. Centre for Ophthalmology, University of Tübingen, Schleichstraße 12–16, 72076 Tübingen, Germany

2. Retina Implant AG, Gerhard-Kindler-Straße 8, 72770 Reutlingen, Germany

3. Mobility Trainer, Mozartweg 11, 72076 Tübingen, Germany

4. Department of Otorhinolaryngology, Division of Paediatric Otorhinolaryngology and Otology, Klinikum Stuttgart, Bismarckstrße 8, 70156 Stuttgart, Germany

5. Department of Ophthalmology, Semmelweis University, Tomo u. 25–29, 1083 Budapest, Hungary

6. Clinic Friedrichstadt, Friedrichstraße 41, 01067 Dresden, Germany

7. MEG Center, University of Tübingen, Otfried-Müller-Straße 47, 72076 Tübingen, Germany

8. STZ EyeTrial at the Centre for Ophthalmology, University of Tübingen, Schleichstraße 12–16, 72076 Tübingen, Germany

Abstract

This study aims at substituting the essential functions of photoreceptors in patients who are blind owing to untreatable forms of hereditary retinal degenerations. A microelectronic neuroprosthetic device, powered via transdermal inductive transmission, carrying 1500 independent microphotodiode-amplifier-electrode elements on a 9 mm 2 chip, was subretinally implanted in nine blind patients. Light perception (8/9), light localization (7/9), motion detection (5/9, angular speed up to 35 deg s −1 ), grating acuity measurement (6/9, up to 3.3 cycles per degree) and visual acuity measurement with Landolt C-rings (2/9) up to Snellen visual acuity of 20/546 (corresponding to decimal 0.037 or corresponding to 1.43 logMAR (minimum angle of resolution)) were restored via the subretinal implant. Additionally, the identification, localization and discrimination of objects improved significantly ( n = 8; p < 0.05 for each subtest) in repeated tests over a nine-month period. Three subjects were able to read letters spontaneously and one subject was able to read letters after training in an alternative-force choice test. Five subjects reported implant-mediated visual perceptions in daily life within a field of 15° of visual angle. Control tests were performed each time with the implant's power source switched off. These data show that subretinal implants can restore visual functions that are useful for daily life.

Publisher

The Royal Society

Subject

General Agricultural and Biological Sciences,General Environmental Science,General Immunology and Microbiology,General Biochemistry, Genetics and Molecular Biology,General Medicine

同舟云学术

1.学者识别学者识别

2.学术分析学术分析

3.人才评估人才评估

"同舟云学术"是以全球学者为主线,采集、加工和组织学术论文而形成的新型学术文献查询和分析系统,可以对全球学者进行文献检索和人才价值评估。用户可以通过关注某些学科领域的顶尖人物而持续追踪该领域的学科进展和研究前沿。经过近期的数据扩容,当前同舟云学术共收录了国内外主流学术期刊6万余种,收集的期刊论文及会议论文总量共计约1.5亿篇,并以每天添加12000余篇中外论文的速度递增。我们也可以为用户提供个性化、定制化的学者数据。欢迎来电咨询!咨询电话:010-8811{复制后删除}0370

www.globalauthorid.com

TOP

Copyright © 2019-2024 北京同舟云网络信息技术有限公司
京公网安备11010802033243号  京ICP备18003416号-3